Manual
9
Note:
1. The users may enter maximum 5 digits.
2. If you enter a current value (incl. digits after the decimal point), and then press ENTER, the PCE-
789 outputs this current value.
3. If the output value is < 0, please, enter the minus first.
4.
If the output value is < 1 and > 0, please, enter “0” first.
Warning:
1. Do not connect any current carrying conductor or a loaded circuit to the SOURCE jacks in order
to avoid the damage of PCE-789.
2. If there is short circuit or an overload at the output jacks, the PCE-789 cannot show any correct
current value. Disconnect the device and check, whether an OUTPUT ERROR symbol is
displayed.
3. Always make only one measurement at the same time and connect only the sockets required for
this. Disconnect all the unused sockets. Connect the measuring cable always only to one socket
type (SOURCE or MEASURE or TC).
b) Setup Instructions
(1) Press SETUP to enter the setup menu.
(2) mA 0%: sets the current starting value of the ramp function. (Details in the chapter "Ramp
function")
(3) mA 100%: sets the current end value of the ramp function. (Details in the chapter "Ramp
function")
(4) 4mA ->: sets the mapping unit for 4mA.
(5) 20mA->: sets the mapping unit for 20mA.
(6)
MAPPING: here the “MAPPING” function can be activated.