Manual
5
Temperature, Thermocouples
(sources and measuring function, 0.1 °C & 0.1 °F resolution, internal cold
junction compensation (CJC), accuracy of the thermocouple not included, 3 minutes after inserting the
thermocouples)
°C
°F
Measuring range
Accuracy
Measuring range
Accuracy
K
-200 to -150
2.0
-382 to -238
3.6
-150 to 0
1.2
-238 to 32
2.1
0 to 1000
0.8
32 to 1832
1.4
1000 to 1370
1.2
1832 to 2498
2.1
J
-200 to -150
2.0
-382 to -238
3.6
-150 to 0
1.0
-238 to 32
1.8
0 to 1050
0.7
32 to 238
1.2
E
-200 to -150
0.9
-382 to -238
2.7
-150 to 0
0.7
-238 to 32
1.6
0 to 850
1.5
32 to 1562
1.2
T
-200 to -150
1.5
-382 to -238
2.7
-150 to 0
1.2
-238 to 32
2.1
0 to 400
0.8
32 to 752
1.4
R
0 to 500
1.8
32 to 932
3.2
500 to 1760
1.5
932 to 3200
2.7
S
0 to 500
1.8
32 to 932
3.2
500 to 1760
1.5
932 to 3200
2.7
N
-200 to 0
1.5
-328 to 32
2.7
0 to 1300
0.9
32 to 2372
1.6
L
-200 to 0
0.9
-328 to 32
1.6
0 to 900
0.7
32 to 1652
1.2
U
-200 to 0
1.1
-328 to 32
1.9
0 to 600
0.7
32 to 1652
1.2
B
600 to 800
2.2
1112 to 1472
3.9
800 to 1000
1.8
1472 to 1832
3.2
1000 to 1820
1.4
1832 to 3308
2.5
C
0 to 1800
1.0
32 to 3272
1.8
1800 to 2310
1.5
3272 to 3308
2.7
mV
-10mV to 70mV
0.05mV
-10mV to 70mV
0.05mV
Voltage (peak-peak for sinusoidal waves)
(Vpp, 0.3 ~ 20kHz, 50% duty cycle, sine wave, 0V offset)
Measuring range
Resolution
Reading accuracy
0.1 to 20V
0.001V
5% ±0.3V
Voltage (peak-peak for non-sinusoidal waves)
(Vpp, 0.3 ~ 20kHz, 0V offset)
Measuring range
Resolution
Reading accuracy
0.1 to 20V
0.001V
6% ±0,4V
Voltage (peak-peak)
(Vpp, 0.3 ~ 20kHz, 50% duty cycle, square wave, 0V offset)
Measuring range
Resolution
Reading accuracy
0.1 to 20V
0.001V
6% ±0,4V
Offset voltage
(max. Vpp < 20V)
Measuring range
Resolution
Reading accuracy
-5 to 5V
0.001V
5% ±0,5V ±5%xVss