![PCE Instruments PCE-789 Manual Download Page 21](http://html1.mh-extra.com/html/pce-instruments/pce-789/pce-789_manual_1516131021.webp)
Manual
21
c) Setup Details
(1) Press S/M to select the desired option.
(2) The desired value can be entered when the appropriate option is in a reverse black / white mode.
(3) C.J.COMP.: here a value for CJC can be set (the standard value is 0.0°C. A compensation
between -5°c and +5°C corresponding to the ideal output values can be set).
(4)
If the option “TC TYPE” has been selected, the desired thermocouple can be selected with the
button SELECT.
(5)
If the option “UNIT” has been selected, °C or ° F can be selected with the help of the button
SELECT.
6
Ramp Function
6.1
Ramp function for voltage sampling (as a source)
a) Setup instructions
(1) Press SETUP to enter the SETUP menu.
(2) V 0%: sets the voltage starting value.
(3) V 100%: sets the voltage end value.
Note:
1. The users may enter maximum 5 digits.
2. If you enter a voltage value (incl. digits after the decimal point), and then press ENTER, the PCE-
789 outputs this voltage value.
3. If the output value is < 0, please, enter the minus first.
4.
If the output value is < 1 and > 0, please, enter “0” first.