background image

 

 

 

 

8161 SERIES DIN RAIL AMPLIFIER MANUAL 

 

10vdc 

5vdc 

1.0

 

Description 

Series  8161  Signal  Conditioner/Amplifier  is  designed  for  use  with  strain  gage  transducers  that 
can operate with a DC excitation-amplification source.  The 8161 series supplies either a 5 volt 
or a 10 volt transducer excitation, which can be selected through an internal jumper (see Figure 
1).    The  8161  series  contains  all  necessary  balancing,  gain,  and  calibration  controls  and 
conditions/amplifies  the  applied  input  to  a  standard  +/-5  or  +/-  10  volt  (jumper  selectable,  see 
Figure 1), and 4-20 mA analog output.  Calibration of the instrument to the transducer is made 
through conventional a shunt technique using an internal calibration resistor.  A front panel push 
button provides for calibration in the positive direction.  
  
2.0

 

Power connection 

The 8161 series is designed to operate on a customer supplied DC power source with a voltage 
range of 

10

 to 28 volts.  Screw terminals 1 and 2 are provided for power connections.  Reference 

Figure 3. 
 
3.0

 

Transducer cabling 

 
High quality twisted pair shielded cable with four conductors is recommended (such as Belden 
8723)  for  transducer  hook  up.    PCB  can  supply  finished  cables  for  transducers  with  either  a 
PT02E-10-6P or  PC04E-10-6P receptacles if desired.  Reference Figure  3 for transducer hook 
up.  

 

 

Figure 1 - Jumper Locations 

P

2

 

P1 

10vdc 

5vdc 

Jumper P1 : Excitation Voltage               

Jumper P2 : Output Voltage 
    

Summary of Contents for 8161-011A

Page 1: ...ge Conditioner Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ... 0 Power Connections 2 3 0 Transducer Cabling 2 4 0 Transducer Excitation 3 5 0 Calibration Resistor 3 6 0 Analog Outputs 3 7 0 Calibration 3 Figures Figure Page 1 Jumper Locations 2 2 SW2 Switch Settings 4 3 Transducer Hookup 6 MANUAL NUMBER 23708 MANUAL REVISION E ECN NUMBER 50103 10 28 19 ...

Page 7: ... to the transducer is made through conventional a shunt technique using an internal calibration resistor A front panel push button provides for calibration in the positive direction 2 0 Power connection The 8161 series is designed to operate on a customer supplied DC power source with a voltage range of 10 to 28 volts Screw terminals 1 and 2 are provided for power connections Reference Figure 3 3 ...

Page 8: ...e to achieve a precisely known equivalent input allowing the instrument sensitivity to be adjusted correctly A 30K ohm resistor is installed at the factory and is used during final checkout of the instrument It should be replaced with the proper value and accuracy for the transducer that it will be used with The resistor is mounted internally on terminals located on the printed circuit board in th...

Page 9: ...time for stabilization of transducer characteristics 4 With the transducer unloaded set the voltage output to ZERO using the coarse and fine zero controls If the current output is to be used set ZERO to 4 mA using the 4 mA adjust 5 Load the transducer to a convenient up scale value which is greater than one half of full scale Adjust the fine span control until the voltage output signal reading is ...

Page 10: ... by the installed calibration resistor 7 If dead weight calibration is not practical and the transducer calibration data is unknown the Equivalent Input value for the factory installed calibration resistor can be approximated as follows if the mV V sensitivity rating of the transducer and the bridge resistance are known X 25000 x Rb K x Rc Where X Equivalent Input of full scale Rb Bridge Resistanc...

Page 11: ...ANUAL 6 1 Ground 2 Vsupply 10 28 VDC 3 Iout 4 20 mA 4 Vout 5 or 10 V 5 Excitation 6 Excitation 7 Signal 8 Signal Figure 3 Transducer Hook up Sig VEXC VSUPPLY 10 28 VDC Ground IOUT VOUT Return VOUT IOUT 4 20 mA VEXC Sig 5 6 7 4 2 8 3 1 ...

Page 12: ...A 2 Physical Size Height x Width x Length 3 35 in x 0 69 in x 2 27 in 89 mm x 18 mm x 58 mm Mounting 35 mm DIN Rail 35 mm DIN Rail All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc OPTIONAL VERSIONS Optional versions have id...

Page 13: ......