MULTICHANNEL SIGNAL CONDITIONER MODEL 482C27 GENERAL OPERATION MANUAL
43
Setting Response: 2:SAVS:ok
Query
:
N/A
Communication Responses
Typically the unit will return
<Unit>:<Cmd String>:OK
when the command is successful. Errors are indicated with
negative numbers. The unit may return one of the following:
<Unit>:<Cmd String>:OK<CR> <LF>
Represents that the last command was entered in the correct format and was performed properly.
<Unit>:<Cmd String>:-1<CR> <LF>
Option Error. The unit is not equipped with the option necessary to implement the command sent
<Unit>:<Cmd String>:-2<CR> <LF>
Channel Error. The channel number in the command is invalid.
<Unit>:<Cmd String>:-3<CR> <LF>
Command Error. The command is not recognized.
<Unit>:<Cmd String>:-4<CR> <LF>
Unit Error. The unit number in the command is invalid.
<Unit>:<Cmd String>:-5<CR> <LF>
Unit Error. The function invoked by the command encountered an error or a query only command (ex.
RBIA) was sent as a setting.
<Unit>:<Cmd String>:-6<CR> <LF>
Command Parameter Error. A channel setting parameter was found to be out of range.