PCB Piezotronics 482C27 General Operations Manual Download Page 21

 

 

 

MULTICHANNEL SIGNAL CONDITIONER MODEL 482C27 GENERAL OPERATION MANUAL

 

13

 

The coupling options are AC or DC. In DC coupling mode the auto zero and auto balance functions become available. 
Auto zero is for all input types and auto balance is for the Differential Voltage and Bridge (full, half and quarter) input 
types. 

 
8) Auto Zero/ Auto Balance (ZERO on front panel):

 

482C27 includes an auto zero function for automatically zeroing channel outputs for any channel in DC coupling mode 
(zeroing  works  for  any  sensor  input  type).  The  auto  balance  function  works  for  channels  that  are  set  to  differential 
voltage, bridge (quarter, full, half), RSE and NRSE input modes ONLY (when DC coupling is enabled). Both zeroing 
functions are not available when a channel is AC coupled. The auto zero function disconnects the input and uses a DAC 
to reduce the channel output to close to 0V. The balance option does the same but leaves the input connected. 
 

EXCi 

EXCv 

FLTo 

ACDC  ZERO  ARNG 

 
When  the  ENTER  button  is  pressed,  the  submenu  for  the  coupling  options  will  include; 

Cancel,  AUTOBAL, 

AUTOZERO

. The selection can be changed by using the up 

 or down 

 buttons. 

 

Chan 

AUTOZERO 

AUTOBAL 

 

NOTE:

 If the zero function fails either the message “Error Removing MainBd DC Offset” or “Error Removing Bridge 

DC Offset” will be displayed depending on the situation. For gains less than 500 the function tries to get below 2mV 
output and gains less than or equal to 500, 50mV is the limit. 

 
9) Auto Range (ARNG on front panel):

 

The auto range (also referred to as auto scale) feature provides an automatic scaling of the output signal. Please refer to 
section 3-6.1 for further description.  With this command, the programmable gain is implemented to give .8 of the full 
scale  output.  The  signals  are  checked  internally  using  the  onboard  A/D.  To  use  this  feature,  place  the  cursor  over 
“ARNG” on the display and press the ENTER key, the following submenu will appear. 
 

EXCi 

EXCv 

FLTo 

ACDC  ZERO  ARNG 

 

Use the up 

 or down 

 arrow keys to toggle between the ‘OFF’, ‘ON’ or ‘IMED’ options. Once you have set the auto 

scale option, hit the ENTER key to return to the main menu. ON turns the auto scale function ON and it will continue 
scaling the gain until it is manually turned off. The ‘IMED’ option runs through the auto scale algorithm once and then 
turns it off automatically. 
 

Summary of Contents for 482C27

Page 1: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Page 2: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Page 3: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Page 4: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Page 5: ...For powering differential voltage MEMS single ended MEMS and bridge quarter full half sensors this signal conditioner provides an effective method for managing small numbers of sensor channels A simp...

Page 6: ...zation 7 The factory defaults are 7 2 3 Main Menu Format 7 2 4 Submenu Format and Command Selections 8 3 0 THEORY OF OPERATION 19 3 1 Sensor Excitation 19 3 2 Input Protection 20 3 3 Input Fault Detec...

Page 7: ...Ethernet Communication 23 4 6 Model 482C27 Communication Guidelines 30 4 7 Model 482C27 Unit Initialization Procedure 30 4 8 Command Summary 30 4 9 Command Format 31 4 10 Commands 32 GAIN 32 SENS 32...

Page 8: ...re 3 Input Fault Window Comparator with LED Indicator 20 Figure 4 Input Amplifier Configuration 20 Figure 5 Autoscale Overload Window Comparator 21 Figure 6 Differential Voltage Full Bridge Type Senso...

Page 9: ...ualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid injury WARNING 2 This equipment is designed with user safety in mind however the protectio...

Page 10: ...d Single Ended Main use is with MEMS sensors that have a single ended output Sensor examples are PCB model series 3711 and 3713 The bridge inputs are compatible with full and bridge sensors Internal s...

Page 11: ...uency 50 or 60 Hz or a multiple of the fundamental frequency is present in the system when the sensors are at rest In order to maintain the operating specification of noise and reduce the effects of l...

Page 12: ...the channel and increments This button is used to scroll UP through the options in the menus In the submenus this button can be used to change the channel and increments This button moves you in the...

Page 13: ...PT from the main menu and hit enter Once in the submenu the flashing cursor will be in the channel column hit the up arrow button until the channel column reads 2 Chan Input Type 2 ICP Hit the left ri...

Page 14: ...le from 1 to 200 0 step of 1 Use the left right arrow key to move the flashing cursor over any digit in the gain column to select an appropriate value In this example a gain of 10 is selected see exam...

Page 15: ...istance If the unit becomes dusty and dirty it may be wiped off with a soft cloth 2 0 FRONT PANEL MODULE CONTROL 2 1 Introduction Model 482C27 is equipped with the capability of controlling functions...

Page 16: ...t Disp Savs GType Pressing the down arrow key button once again shows GAIN SENS INPT FSOT FSIN FLTi 2 4 Submenu Format and Command Selections 1 Gain Functionality Includes five menu selections The fol...

Page 17: ...GType must be set to the Normalize selection The submenus for SENS FSIN and FSOT appear like this The 482C27 has the ability to automatically normalize the output sensitivity of each channel based on...

Page 18: ...to select either On or Off once you have the desired setting selected hit ENTER to apply it Chan INP Filter 2 Off 4 Output Filter FLTo on front panel A screen stating This Option is Not Installed will...

Page 19: ...low it to be set while unless it is in ICP mode If a channel is selected from the EXCi menu that is inappropriate then NA will appear as shown below in place of the ICP current settings Also because o...

Page 20: ...will not allow it to be set while in those modes If a channel is selected from the EXCv menu that is inappropriate then NA will appear as shown below in place of the voltage and polarity settings Als...

Page 21: ...AUTOBAL AUTOZERO The selection can be changed by using the up or down buttons Chan AUTOZERO 1 AUTOBAL NOTE If the zero function fails either the message Error Removing MainBd DC Offset or Error Removi...

Page 22: ...7 A This Option is Not Installed message will appear if it is selected SWOT CLMP Bias CHRD Cal LEDS 12 Bias Measurement Bias on the front panel The 482C27 is capable of measuring the bias voltage pres...

Page 23: ...g the ENTER key on the front panel when the cursor is located on the CAL option CHRD CAL LEDS Ver Unit Reset The following submenu will appear Choose the channel you wish to change by using the up and...

Page 24: ...ion Ver on front panel To determine the firmware version of the unit select the Ver menu option CHRD Cal LEDS Ver Unit Reset A screen similar to the one shown below will be displayed PCB 482C27 FW v4P...

Page 25: ...ay Content Options Disp on front panel The display option allows the user to choose from 3 selections of front panel content To enter this mode place the cursor over the Disp option and hit enter Ver...

Page 26: ...eplaces the gain setting The bias reading will be refreshed periodically Following is what the screen will look like if the display was set to show Bias B B I I B15 B15 B23 B23 The Output display opti...

Page 27: ...w impedance voltage output The sensor is powered by a 24 VDC power supply having a typical constant current of 2 to 20 mA current can be fixed or variable depending on the signal conditioner model Sen...

Page 28: ...front panel input fault LED lights NOTE Red LED implies input fault Figure 3 Input Fault Window Comparator with LED Indicator 3 4 Input Interface The input signal conditioning for each channel provid...

Page 29: ...justment appears as ARNG on front panel display It first sets maximum gain on all channels then decreases the gain setting of any channel on which an overload has occurred Auto scale continues until t...

Page 30: ...ferential voltage and bridge quarter half full mode sensor inputs NOTE If the sense lines are not used they should be tied to the excitation lines 3 10 RS 232 Ethernet connections See section 4 0 for...

Page 31: ...umber of stop bits 1 Handshaking None Echo None Transmit EOL Carriage Return and Line Feed CR LF Receive EOL Carriage Return and Line Feed CR LF 4 4 RS 232 Rear Panel Pinout Listing Number of 9 Pin Co...

Page 32: ...in the list You can verify it is the correct unit but comparing the Hardware MAC Address displayed to the one listed on the side of the 482 unit Selecting an item from the list shown below highlights...

Page 33: ...rs though are not the default parameters of this device In the screens that follow the fields pointed to with arrows are the modified parameters Should you need to change them or want to check them if...

Page 34: ...MANUAL 26 After the Login dialog the Settings pane will appear on the left as shown below click on Connection Make sure the parameters are set as shown If you need to change the Port do it here When...

Page 35: ...MULTICHANNEL SIGNAL CONDITIONER MODEL 482C27 GENERAL OPERATION MANUAL 27...

Page 36: ...MULTICHANNEL SIGNAL CONDITIONER MODEL 482C27 GENERAL OPERATION MANUAL 28...

Page 37: ...MULTICHANNEL SIGNAL CONDITIONER MODEL 482C27 GENERAL OPERATION MANUAL 29...

Page 38: ...el LEDs three times 4 8 Command Summary The table below is a summary of the 482C27 command set The 482C series is highly differentiated and some commands may not be valid in all units The 482C command...

Page 39: ...onse being returned No response is ever given to a Global Unit command All messages must be terminated with a CR r and LF n combination Command Format Unit Ch Cmd value1 value2 Ch Cmd value1 value2 n...

Page 40: ...0 10 0 200 0 3 5 0 10 0 10 0 200 0 4 5 0 10 0 10 0 200 0 SENS The SENS command provides a mechanism to have the transducer Sensitivity influence the Gain setting of the channel Channel Gain is calcul...

Page 41: ...ong with the transducer sensitivity set the gain Channel Gain is calculated using the equation Gain FSO 1000 FSI SENS Setting 1 1 FSCO 10 000 r n unit 1 channel 1 FSO set to 10 0 Setting Response 1 FS...

Page 42: ...P Voltage mode maximum is 200 so switching from Bridge to ICP Voltage will reset the Gain to the ICP maximum if the current Bridge setting is higher than the ICP maximum IEXC The IEXC command sets the...

Page 43: ...Filter The Input Filter value is sent as an integer value of either 0 Disable or 1 Enable Setting 2 1 FLTR 1 r n unit 2 channel 1 Input Filter Enabled Setting Response 2 FLTR ok Query The FLTR query...

Page 44: ...1 1 CPLG 1 r n unit 1 channel 1 DC Coupled Setting Response 2 CPLG ok Query The CPLG query returns all channels coupling setting Query Format Unit Ch CMD Response format Unit Cmd Ch 0 1 Query 1 1 CPLG...

Page 45: ...ration setting Query Format Unit Ch CMD Response format Unit Cmd Ch 0 1 Query 1 1 CALB Response 1 CALB 1 4 Global Query 1 0 CALB Global Response 1 CALB 1 4 2 0 3 0 4 0 RTED The Read TEDS RTED command...

Page 46: ...nly Query This command must be a directed command use of the global channel indicator is not allowed Query Format Unit Ch CMD Response format Unit Cmd Ch GAIN Gain value SENS SENS value FSCI FSCI valu...

Page 47: ...nnel settings on power up Unit Bit 1 1 BAD EEPROM read for Unit options on power up Unit Bit 2 1 BAD EEPROM read for cal factors on power up For Unit bit map 0 no errors Channel Bit 0 Short input faul...

Page 48: ...ILTER_PGMBTR 0x10 Prgm LP Butterworth filter Miscellaneous Options OPT_MISC_COUPLING 0x01 AC DC coupling OPT_MISC_CLAMP 0x02 Clamp OPT_MISC_TEDS 0x04 TEDS OPT_MISC_IEXC 0x08 current excitation OPT_MIS...

Page 49: ...e following error codes could be reported by the auto zero balance function 11 Bridge DC Offset ERR Illegal Setting 12 Bridge DC Offset ERR Too Many iterations 13 ICP DC Offset ERR Bad RDG 14 ICP DC O...

Page 50: ...e and then turns off the function automatically When Auto scale is ON the unit will continue to invoke the function adjusting the gain for the current input level until the function is turned OFF with...

Page 51: ...d properly Unit Cmd String 1 CR LF Option Error The unit is not equipped with the option necessary to implement the command sent Unit Cmd String 2 CR LF Channel Error The channel number in the command...

Page 52: ...lines must be connected The figures below illustrate the connections in each of these modes NOTE The shell of the DIN connector is connected as shown to earth ground internally 5 1 Differential Voltag...

Page 53: ...e signal conditioner 5 2 Half Bridge Mode Figure 8 Recommended Connections for Half Bridge Type Sensors sense leads wired for optimum performance full bridge sensor SIG GND optional shunt resistor ext...

Page 54: ...d be set to RSE mode which internally grounds the SIG input The voltage output from the sensor or other source may then be applied to the SIG non inverting input Set V EXC to unipolar at a voltage app...

Page 55: ...rol keypad 1 Input Mode Select INPT on front panel Set sensor input type to Differential Voltage 2 Gain Type GType on front panel Set the gain type to GAIN 3 Gain GAIN on front panel Set signal condit...

Page 56: ...ioner channel gain to 1 on all channels connected to a MEMS sensor 4 Coupling ACDC on front panel Set signal conditioner input coupling to DC 5 Voltage Excitation EXCv on front panel Set the excitatio...

Page 57: ...10 Hz Gain x1 1 5 V Hz 1 5 V Hz 3 Spectral Noise 100 Hz Gain x1 1 0 V Hz 1 0 V Hz 3 Spectral Noise 1 kHz Gain x1 1 0 V Hz 1 0 V Hz 3 Spectral Noise 10 kHz Gain x1 1 0 V Hz 1 0 V Hz 3 Broadband Electri...

Page 58: ......