background image

 

 

 

MODEL 482B06 / F482B06 LINE POWERED SIGNAL CONDITIONER

 

4

 

expressly  specified.  Damage  to  instruments  caused  by 
incorrect  power  or  misapplication  is  not  covered  by 
warranty.  If  there  are  any  questions  regarding  power, 
intended  application,  or  general  usage,  please  contact  a 
PCB application engineer (or your local sales contact). 

 

6.0 Maintenance and Repair 

 
Because 

of 

the 

sophisticated 

nature 

of 

PCB 

instrumentation,  field  repair  of  the  equipment  is  not 
recommended.  Most  PCB  sensors  are  of  modular 
construction  and  are  factory  repairable.  A  repair  or 
replacement  quotation  is  available  at  no  charge.  Before 
returning  equipment  for  repair,  it  is  strongly  suggested 
that  the  user  confer  with  a  factory  application  engineer 
(or  local  sales  contact)  concerning  the  difficulty  and  to 
ascertain if an on-site procedure rectifies the problem. 
 
If  repair  is  indicated,  contact  PCB  to  request  a  Return 
Materials  Authorization  (RMA)  number  from  the 
factory.  An  advanced  authorization  to  proceed  with  the 
repair,  permitting  charges  of  up  to  50%  of  a  new  item, 
greatly expedites repair. 
 
For the most efficient service, please provide a detailed, 
written  description  of  the  malfunction  encountered  with 
the  returned  equipment,  together  with  your  specific 
application and setup procedures. Customers outside the 
U.S.  should  consult  their  local  sales  contact  for 
information  on  returning  equipment.  For  exceptions, 
please  contact  the  International  Sales  Department  at 
PCB  to  request  shipping  instructions  and  an  RMA.  For 
assistance, please call (716) 684-0001, or fax us at (716) 
684-0987. You may also receive assistance via   email at 

[email protected]

 or our web site at 

www.pcb.com. 

 

MANUAL NUMBER:  19181 
MANUAL REVISION:  NR 

Summary of Contents for 482B06

Page 1: ...oner for ICP Sensors Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ... repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment calibration cycles are typically established by the users own quality regimen When in doubt about a calibration cycle a good rule...

Page 3: ...ory customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and mounting hardware are not covered by warranty Mechanical damage...

Page 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Page 6: ...duction 1 2 0 Description 1 3 0 Installation 2 4 0 Coupling Time Constant AC Coupled 2 4 1 Driving Long Cables 3 4 2 Setting the Constant Current 3 5 0 Warranty 3 6 0 Maintenance and Repair 4 Appendix Specification Sheet Outline Drawing Table of Figures Figure 1 482B06 Front Panel 2 Figure 2 Fault Monitor Table 2 Figure 3 Constant Current Adjust 3 ...

Page 7: ...ld therefore be done ONLY at an ESD safe work area Many products have ESD protection but the level of protection may be exceeded by extremely high voltage The Caution heading used in this manual explains hazards that could damage the instrument 1 2 Introduction The Model 482B06 F482B06 is a single channel line operated signal conditioner for use with ICP sensor This unit provides a 24VDC 4mA const...

Page 8: ...ect a typical bias of 8 to 12 volts It does not read midscale if the transducer bias is 3 to 5 volts and may be closer to red which is expected for low bias sensors If a cable is faulty open or the sensor s built in amplifier is open the meter indicates a yellow full scale reading When a cable or sensor is shorted the meter indicates zero volts red Connect the output jack to a read out instrument ...

Page 9: ... Another technique requires the use of the Model 073A Impedance Matching Resistor connected immediately after the sensor This variable resistor is then used to tune the current to the line One of the best approaches is to test the long line with a signal generator set at a frequency of interest 4 2 Setting the Constant Current To set the constant current locate the current adjust potentiometer on ...

Page 10: ...oncerning the difficulty and to ascertain if an on site procedure rectifies the problem If repair is indicated contact PCB to request a Return Materials Authorization RMA number from the factory An advanced authorization to proceed with the repair permitting charges of up to 50 of a new item greatly expedites repair For the most efficient service please provide a detailed written description of th...

Page 11: ... BNC Jack Connector Output BNC Jack BNC Jack Connector AC Power Input IEC 320 IEC 320 Size Height x Length x Width 4 3 in x 6 0 in x 1 8 in 10 92 cm x 15 24 cm x 4 57 cm Weight 1 2 lb 544 gm All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark ...

Page 12: ......