background image

1

1

2

2

3

3

4

4

5

5

6

6

7

7

8

8

A

A

B

B

C

C

D

D

DWG. NO.

SCALE:

SHEET

DRAWN

CHECKED

ENGINEER

TITLE

UNLESS OTHERWISE SPECIFIED TOLERANCES ARE:

DIMENSIONS IN MILLIMETERS

[ IN BRACKETS ]

ANGLES

  

`

 .5 DEGREES

24350 Indoplex Circle, Farmington Hills,MI 48335
(716)  684-0001 E-MAIL:  [email protected]

DIMENSIONS  IN INCHES

ANGLES

  

`

 

.5 DEGREES

FILLETS AND RADII 

.015 MAX 

FILLETS AND RADII 

0.38 MAX

OUTLINE DRAWING

49605

1 OF 1

FULL

1203/1403 SERIES, PT CONNECTOR

LOW PROFILE LOAD CELLS

PTE

11/21/11

DECIMALS     XX   ±.01

XXX   ±.005

DECIMALS    X   ± 0.3

XX   ± 0.13

A

B

B

A

4

9

6

0

5

PCB Load & Torque Inc. claims proprietary rights  in 
the  information disclosed hereon.  Neither it nor any 
reproduction thereof will be disclosed to others without
the written consent of PCB Load and Torque Inc.

1.37 [34.8]

.12 [3.0]

n

1.34 [34.0]

n

4.12 [104.6]

.62 [15.7]

"THREAD" 

x

.88 [22.3]

(STANDARD LOAD CELLS 5/8-18 UNF-3B,

METRIC "M" LOAD CELLS M16 X 2.0 4H,

SPECIAL LOAD CELL 1"-14UNS-3B)

 

n

3.500 [88.90]

n

4.12 [104.6]

n

1.34 [34.0]

1.13 [28.7]

.5 [13]

n

.25 [6.4]

4X 

 

x

.25 [6.4]

     (SPANNER WRENCH)

CONNECTOR "A"
PT02E-10-6P CONNECTOR
WIRING CODE: TYPICAL
PIN A +EXC
PIN B +SIG
PIN C -SIG
PIN D -EXC
PIN E N/C
PIN F N/C
 

n

3.500 [88.90]

22.5°

8X 1/4-28 UNF-2B 

x

.50 [12.7]

"THREAD" 

x

.88 [22.3]

(STANDARD LOAD CELLS 5/8-18UNF-3B,

METRIC "M" LOAD CELLS M16X2.0 4H,

SPECIAL LOAD CELL 1"-14UNS-3B)

2.50 [63.5]

.030 [.76]

45°

"THREAD" 

x

.88 [22.3]

(STANDARD LOAD CELLS 5/8-18 UNF-3B,

METRIC "M" LOAD CELLS M16 X 2.0 4H,

SPECIAL LOAD CELL 1"-14UNS-3B)

2.67 [67.8]

2.54 [64.6]

 W/CONN. PROTECTOR

n

1.34 [34.0]

.12 [3.0]

CONNECTOR "A"
PT02E-10-6P CONNECTOR
WIRING CODE: TYPICAL
PIN A +EXC
PIN B +SIG
PIN C -SIG
PIN D -EXC
PIN E N/C
PIN F N/C
 

.030 [.76]

"THREAD" 

x

.88 [22.3]

(STANDARD LOAD CELLS 5/8-18UNF-3B,

METRIC "M" LOAD CELLS M16X2.0 4H,

SPECIAL LOAD CELL 1"-14UNS-3B)

n

4.12 [104.6]

45°

2.67 [67.8]

2.54 [64.6]

 W/CONN. PROTECTOR

45°

n

.281 [7.14]

8X 

 THRU

22.5°

8X 1/4-28 X 1.75 [44.5] 12 POINT FLANGE BOLT

CONNECTOR "B"
DUAL BRIDGE ONLY

REVISIONS

REV

DESCRIPTION

ECO

E

HUB DIAMETER WAS 1.25 [31.8] - 10.05.16, PTE

46027

1

1

SUPPLIED WITH MATING FASTENERS.

1203, M1203, 1403 AND M1403 SERIES LOAD CELLS

INCLUDING DUAL BRIDGE

1203, M1203, 1403 AND M1403 SERIES LOAD CELLS W/BASES

INCLUDING DUAL BRIDGE

1203, M1203, 1403 AND M1403 SERIES BASES

n

1.34 [34.0]

MDF

11/21/11

JDM

11/21/11

WIRING CODE FOR "T" SERIES LOAD CELLS
 
     PIN A   +EXC.
     PIN B   +SIG.
     PIN C   -SIG.
     PIN D   -EXC.
     PIN E    TEDS DATA
     PIN F    TEDS GROUND

2

2

2

3

3

MODEL 084A90 CONNECTOR PROTECTOR SHOWN:
OPTIONAL ON ALL 1203/M1203 SERIES AND 1403/M1403 SINGLE BRIDGE LOAD CELLS
STANDARD ON 1403/M1403 SERIES DUAL BRIDGE LOAD CELLS

3

.21 [5.4]

.21 [5.4]

CONNECTOR "B"
DUAL BRIDGE ONLY

.25 [6.4]

.62 [15.7]

4X Ø.25 [6.4] 

x

.25 [6.4]

    (SPANNER WRENCH)

(

n

1.34 [34.0])

Summary of Contents for 1203 Series

Page 1: ...OFILE LOAD CELL Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ...r 4 5 0 ELECTRICAL INSTALLATION 5 5 1 Electrical Drawing Western Regional Std 5 5 2 Cable Grounding Considerations 5 6 0 CALIBRATION 5 6 1 Calibration Certificate Description 5 6 1 1 Measured Output 6 6 1 2 Hysteresis 6 6 1 3 Best Fit Output 6 6 1 4 Strain Gage Measurements 6 6 1 5 Shunt Calibration Standard Resistor 6 6 1 6 Static Error Band SEB 6 7 0 MOMENT COMPENSATION 6 8 0 FATIGUE OVERLOAD 6 ...

Page 7: ...evaluated for potential failure modes of the fixtures It is important to use threaded fasteners properly when assembling test fixtures used with the load cells Fixtures with inadequately designed bolted joints or improperly tightened threaded fasteners can fail before the parts under test have completed the planned fatigue test cycles 3 0 OVERVIEW General purpose low profile load cells come in a v...

Page 8: ...r modified 4 2 Mounting Load Cell to a Standard Base or Custom Fixture If the load cell is to be mounted to a custom base or fixture it is necessary to consider the following 1 The thickness of the new base or fixture must not be any thinner than the original base supplied with the load cell Materials used must have the same temperature coefficient of expansion for example 4140 Steel 2 The mountin...

Page 9: ...of Connector Protector 1 Remove the 4 40 x 1 4 button head cap screws from the connector using a 1 16 hex wrench as shown in Figure 7 2 Carefully pull the connector from the load cell Take care to not break or loosen any wire connections 3 Slide the connector protector 54823 02 behind the connector as shown in Figure 8 4 Apply a drop Loctite Threadlocker Blue 242 to the threads of the 4 40 x 3 8 s...

Page 10: ... A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded If the power cord ground and structure ground are not at the same potential it may be necessary to provide a secure structure ground perhaps by driving a copper rod and attaching a ground strap 6 0 CALIBRATION Every ge...

Page 11: ...s into the load cell structure The strain gage bridges on general purpose low profile load cells are adjusted such that the error due to eccentric loading is less than 0 25 per inch of offset for 1200 series load cells and 0 1 per inch of offset for fatigue rated 1400 series load cells Side loads moments and torque must be limited to avoid electrical or mechanical damage to the load cell The allow...

Page 12: ...rces on the load cell If possible remove all loads 2 Power up the host signal conditioner and connect it to the load cell via appropriate cable and allow for a 30 minute warm up 3 Set the load indicator display to read exactly 00 000 4 Connect the shunt resistor to the terminals specified in the calibration certificate and adjust the span or gain until the display reads the force value stated on t...

Page 13: ...he load cell is mounted on 5 Check the signal conditioning electronics for proper setup 6 Check the insulation resistance of shielded conductors for short circuits 7 Check isolation resistance load cell flexure to conductors 8 Check load cell bridge resistances A D excitation and B C the signal leads 9 Check bridge balance 10 Keep a record of your observations correct problems or contact PCB facto...

Page 14: ...rder Please request a return material authorization RMA before sending a load cell back to the factory for any reason For calibration services if possible a copy of the purchase order covering the requested services should be included with the returned load cell 13 0 WARRANTY Standard warranty on general purpose low profile load cells covers parts and workmanship For full details refer to the Warr...

Page 15: ...nted Steel Sensing Element Strain Gage Strain Gage Mounting Torque 200 lbf in 23 Nm Deflection at Full Scale Capacity 0 0005 in 0 013 mm Electrical Connector PT02E 10 6P PT02E 10 6P All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice 1403 12B PC04E 10 6P Connector 1403 12C...

Page 16: ... LOAD CELLS 5 8 18UNF 3B METRIC M LOAD CELLS M16X2 0 4H SPECIAL LOAD CELL 1 14UNS 3B 2 50 63 5 030 76 45 THREAD x 88 22 3 STANDARD LOAD CELLS 5 8 18 UNF 3B METRIC M LOAD CELLS M16 X 2 0 4H SPECIAL LOAD CELL 1 14UNS 3B 2 67 67 8 2 54 64 6 W CONN PROTECTOR n1 34 34 0 12 3 0 CONNECTOR A PT02E 10 6P CONNECTOR WIRING CODE TYPICAL PIN A EXC PIN B SIG PIN C SIG PIN D EXC PIN E N C PIN F N C 030 76 THREAD...

Reviews: