GENERAL PURPOSE LOW PROFILE LOAD CELL OPERATION MANUAL 

 

5

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

5.0 ELECTRICAL INSTALLATION 

 
 

 
 
 
 
 
 
 
 
 

5.1 Electrical Drawing / Western Regional Std. 

All  load  cells  are  wired  following  the  Western  Region 
Standard.    All  models  utilize  strain  gages  configured  into  a 
Wheatstone  Bridge  Circuit  to  produce  the  primary  sensing 
element.  The  four-arm  Wheatstone  bridge  configuration  is 
shown below in Figure 11. 

 
 
 
 
 
 
 
 
 

The  gages  are  bonded  to  the  load  cell's  structure. 
Typically,  a  regulated  DC  or  AC  excitation  is  applied 
between A and D of the bridge. When a force is applied to 
the  load cell,  the Wheatstone  bridge  becomes  unbalanced, 
causing  an  output  voltage  between  B  and  C,  which  is 
proportional to the applied load.  This configuration allows 
for  temperature  and  pressure  compensation,  as  well  as 
cancellation  of  signals  caused  by  forces  not  directly 
applied to the axis of the applied load. Output is typically 
expressed in units of mV/V of excitation. 
 

5.2 Cable & Grounding Considerations 

Proper  grounding  and  shielding  is  required  to  prevent 
electrical  noise  in  strain  gage  load  cell  measuring  systems.  
The cable must be shielded twisted pairs with a drain wire. 
 
Cable shields must be grounded only at one end, for example, 
on  the  instrument  or  control  system  ground.    The  load  cell 
case is grounded by mechanical attachment to the structure to 
which it is mounted. 
 
The  instrument  or  control  system  is  grounded  through  its 
power cord.  Ground loops and measuring system wiring may 
result in unstable or noisy signals. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
A  simple  test  with  a  voltmeter  connected  between  the  power 
cord  ground  and  the  structure  on  which  the  load  cell  is 
mounted  can  confirm  that  the  structure  has  been  properly 
grounded.  If the power cord ground and structure ground are 
not  at  the  same  potential,  it  may  be  necessary  to  provide  a 
secure  structure  ground,  perhaps  by  driving  a  copper  rod  and 
attaching a ground strap. 
 
 

6.0 CALIBRATION 

 
Every  general  purpose  low  profile  load  cell  purchased  from 
PCB Load & Torque, Inc. has been fully calibrated in tension 
and compression per ISO/IEC 17025 procedures, and meets all 
published  specifications.    Each  load  cell  will  come  with  a 
calibration  certificate  designated  with  matching  model  and 
serial  numbers.  PCB  Load  &  Torque  also  offers  calibration 
services on an on-going basis. 
 

6.1 Calibration Certificate Description 

Calibration reports supplied with PCB Load & Torque general 
purpose low profile load cells contain valuable information to 
assist  the  customer  in  use  of  the  equipment.  A  separate 
calibration  report  is  provided  for  tension  and  compression 
calibrations 

on 

each 

bridge. 

Calibration 

procedures, 

equipment, and reports comply with ISO/IEC 17025. 

Load Cell Receptacle “A”: 

PT02E-10-6P 

Mating Connector: 

PT06A-10-6S 

PCB Item No: 

181-012A 

Load Cell Receptacle “B”: 

PC04E-10-6P 

Mating Connector: 

PC06A-10-6S 

PCB Item No: 

182-025A 

Load Cell Receptacle “C”: 

PT02E-12-8P 

Load Cell Receptacle “D”: 

PC01E-12-8P 

 

Table 6

 

– Electrical Connection Options 

Figure 12

 

- Grounding 

 

Test Rig Ground 

Power Supply 

Instrument Ground 

Cable Shielding 

Load Cell 

 

- Signal White 

- Excitation Black 

+ Excitation Red 

+ Signal Green 

Figure 11 - Western Regional Strain Gage Committee Wiring Code 

Figure 9 - Attach the Connector 

Protector (54823-02) using #4-40 x 

3/8” Socket Head Cap Screws 

Figure 10 - Fully Assembled 

 

 

Summary of Contents for 1203 Series

Page 1: ...OFILE LOAD CELL Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ...r 4 5 0 ELECTRICAL INSTALLATION 5 5 1 Electrical Drawing Western Regional Std 5 5 2 Cable Grounding Considerations 5 6 0 CALIBRATION 5 6 1 Calibration Certificate Description 5 6 1 1 Measured Output 6 6 1 2 Hysteresis 6 6 1 3 Best Fit Output 6 6 1 4 Strain Gage Measurements 6 6 1 5 Shunt Calibration Standard Resistor 6 6 1 6 Static Error Band SEB 6 7 0 MOMENT COMPENSATION 6 8 0 FATIGUE OVERLOAD 6 ...

Page 7: ...evaluated for potential failure modes of the fixtures It is important to use threaded fasteners properly when assembling test fixtures used with the load cells Fixtures with inadequately designed bolted joints or improperly tightened threaded fasteners can fail before the parts under test have completed the planned fatigue test cycles 3 0 OVERVIEW General purpose low profile load cells come in a v...

Page 8: ...r modified 4 2 Mounting Load Cell to a Standard Base or Custom Fixture If the load cell is to be mounted to a custom base or fixture it is necessary to consider the following 1 The thickness of the new base or fixture must not be any thinner than the original base supplied with the load cell Materials used must have the same temperature coefficient of expansion for example 4140 Steel 2 The mountin...

Page 9: ...of Connector Protector 1 Remove the 4 40 x 1 4 button head cap screws from the connector using a 1 16 hex wrench as shown in Figure 7 2 Carefully pull the connector from the load cell Take care to not break or loosen any wire connections 3 Slide the connector protector 54823 02 behind the connector as shown in Figure 8 4 Apply a drop Loctite Threadlocker Blue 242 to the threads of the 4 40 x 3 8 s...

Page 10: ... A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded If the power cord ground and structure ground are not at the same potential it may be necessary to provide a secure structure ground perhaps by driving a copper rod and attaching a ground strap 6 0 CALIBRATION Every ge...

Page 11: ...s into the load cell structure The strain gage bridges on general purpose low profile load cells are adjusted such that the error due to eccentric loading is less than 0 25 per inch of offset for 1200 series load cells and 0 1 per inch of offset for fatigue rated 1400 series load cells Side loads moments and torque must be limited to avoid electrical or mechanical damage to the load cell The allow...

Page 12: ...rces on the load cell If possible remove all loads 2 Power up the host signal conditioner and connect it to the load cell via appropriate cable and allow for a 30 minute warm up 3 Set the load indicator display to read exactly 00 000 4 Connect the shunt resistor to the terminals specified in the calibration certificate and adjust the span or gain until the display reads the force value stated on t...

Page 13: ...he load cell is mounted on 5 Check the signal conditioning electronics for proper setup 6 Check the insulation resistance of shielded conductors for short circuits 7 Check isolation resistance load cell flexure to conductors 8 Check load cell bridge resistances A D excitation and B C the signal leads 9 Check bridge balance 10 Keep a record of your observations correct problems or contact PCB facto...

Page 14: ...rder Please request a return material authorization RMA before sending a load cell back to the factory for any reason For calibration services if possible a copy of the purchase order covering the requested services should be included with the returned load cell 13 0 WARRANTY Standard warranty on general purpose low profile load cells covers parts and workmanship For full details refer to the Warr...

Page 15: ...nted Steel Sensing Element Strain Gage Strain Gage Mounting Torque 200 lbf in 23 Nm Deflection at Full Scale Capacity 0 0005 in 0 013 mm Electrical Connector PT02E 10 6P PT02E 10 6P All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice 1403 12B PC04E 10 6P Connector 1403 12C...

Page 16: ... LOAD CELLS 5 8 18UNF 3B METRIC M LOAD CELLS M16X2 0 4H SPECIAL LOAD CELL 1 14UNS 3B 2 50 63 5 030 76 45 THREAD x 88 22 3 STANDARD LOAD CELLS 5 8 18 UNF 3B METRIC M LOAD CELLS M16 X 2 0 4H SPECIAL LOAD CELL 1 14UNS 3B 2 67 67 8 2 54 64 6 W CONN PROTECTOR n1 34 34 0 12 3 0 CONNECTOR A PT02E 10 6P CONNECTOR WIRING CODE TYPICAL PIN A EXC PIN B SIG PIN C SIG PIN D EXC PIN E N C PIN F N C 030 76 THREAD...

Reviews: