Page 56
WEIGHING PARAMETERS
The parameters allow to adjust the acquisition and updating times of the display and the manual or
automatic resetting of the instrument.
STABILITY FACTOR [1303]
This parameter defines the number of divisions necessary to consider the weight as stable. A large
number of divisions allows the instrument to quickly detect weight stability, which is necessary when
performing tare and printing commands.
N.B. If the value selected is 0, the value 0 must be set in the TIMEOUT parameter.
Selectable value
0÷4
Default: 2
AUTOZERO SETPOINT [1304-1305]
This parameter defines the value of the maximum weight that can be reset at power up.
This operation corresponds to a zero calibration of the system and is only performed if the weight is
stable and lower than the set value.
Value from 0 to Cell capacity.
Default: 0
ZERO TRACKING [1306]
The function allows to perform a momentary zero calibration, thus compensating for any thermal drift
of the weight. When the instrument is turned off, the previous Zero calibration is automatically restored.
The maximum weight that can be zeroed from this parameter is 2% of the system capacity. To disable
the function, set the value 0.
Value
Change
0
Check excluded
1
0.5 div/sec
2
1 div/sec
3
2 div/sec
4
3 div/sec
Default: 0
ZERO BAND [1307-1308]
This parameter defines the number of divisions that can be reset from the touchscreen or via the asso-
ciated input.
Value from 0 to 200 divisions.
Default: 100