Configuration
The ASLD 2200 requires very little configuration, the user must check the following details
before use. If you find that the incorrect sensor or encoder disc has been purchased contact
Partech to arrange a replacement.
Sensor Selection
The Soli-Tech 20 sensor is available in four ranges; which are shown below along with typical
applications.
Part Number
Description
Typical Application
179100
Soli-Tech 20, range 0-200
IR100, range 0-200
Water Treatment Clarifiers
179120
Soli-Tech 20, range 0-1500
IR40, range 0-1500
Sewage Treatment Final Settlement
Water Treatment Primary
179140
Soli-Tech 20, range 0-10000
IR15, range 0-10000
Sewage Treatment Primary
Water Treatment Thickeners
179160
Soli-Tech 20, range 0-30000
IR8, range, 0-30000
Sewage Treatment Thickeners
Depth Range Selection
The depth range depends on the encoder disc that has been fitted to the instrument, this can be
found near the motor assembly.
Motor Assembly Version 1
(Thin stainless steel side plates)
Motor Assembly Version 2
(Thick aluminium side plates)
No of Blades
Minimum Range
(metres)
Maximum Range
(metres)
Minimum Range
(metres)
Maximum Range
(metres)
1
5.6
13.8
11.0
20.0
2
2.8
6.9
5.5
11.0
4
1.4
3.4
3.0
6.0
6
1.0
2.2
2.0
4.0
8
0.8
1.6
1.5
2.5
The instrument will have been fitted with one of these encoder discs, and will have been
selected prior to order placement. The encoder disc can be changed in the field if required, see
the rear of the manual for part numbers.