ISO Explanation
The ISO 4406:1999 standard currently
reflects the number of particles greater
than 4 microns, 6 microns and 14 microns.
The 4 and 6 micron ranges are used to give
an indication of the "silting" condition of a
fluid, while the 14 micron range reflects the
presence of larger "wear catalysts".
15
13
09
>2
>5
>15
160 - 320
40 - 80
2.5 - 5
Range Number
Micron
Actual Particle Count Range (per ml)
Particles
> 2 microns
Particles
> 5 microns
Particles
> 15 microns
ISO CODE
15 / 13 / 09
The ISO classification shown below is
15/13/09. These values are range numbers
that correlate to the ISO 4406 chart
(Figure 6).
If your IQ-200 reports to ISO 4406:1987 modified,
the data would be broken down as follows:
Particles
> 4(c) microns
Particles
> 6(c) microns
Particles
> 14(c) microns
ISO CODE
15 / 13 / 09
If your IQ-200 reports to the new ISO 4406:1999,
the data would be broken down as follows:
15
13
09
>4(c)
>6(c)
>14(c)
160 - 320
40 - 80
2.5 - 5
Range Number
Micron
Actual Particle Count Range (per ml)
Data Interpretation