Sample
number
Volume
per
Sample
(ml)
AMF* (ml)
Approx.
Actual Test
Sample
Volume (ml)
1
60
5
75
2
5 *
60
5
70
3
5 *
60
-
65
Total
180
10
210
10
Flush
Volume
(ml)
20
* A utomatic M ini- F lush
Example B:
The icountBS unit has been configured
– Flush Volume = 20ml, No of Samples = 3, Volume per sample = 50ml
Sample
number
Volume
per
Sample
(ml)
AMF* (ml)
Approx
Actual Test
Sample
Volume (ml)
1
50
5
75
2
5 *
50+50
-
105
3
-
-
-
Total
150
5
180
20
Flush
Volume
(ml)
25
* A utomatic M ini- F lush
-
I) Review New Test Data
On completion, the test result is shown.
a.
Select “Show Counts” to review test counts.
b.
On the test counts screen, if the test has more than one measured sample, then press/select the
and
to
page through the total and each sample.
c.
Select “Show Bottle” to review test bottle detail (return back to the original screen).
The test data is automatically stored on the icountBS (refer to section K for instructions to browse test data).
The icountBS is limited to 500 stored tests. By default, the icountBS will warn the user from the 450
th
test to export or
delete test data. At the 500
th
test, the user has no further option but to export / delete test data. Refer to
section N
of this
manual to change this setting.
J) Perform Another Test
Straight after review the test data it is possible to perform another test
– assuming that there is plenty of fluid left in the
sample bottle.
Test bottle screen
a
c
b
Test bottle screen
Test bottle counts screen
Test bottle screen
Summary of Contents for icountBS
Page 1: ...B 89 809 Rev C Parker Hannifin 2015 www parker com hfde icountBS User Manual GB ...
Page 30: ...B 89 809 Rev C 30 Parker Hannifin 2015 www parker com hfde ...
Page 31: ...B 89 809 Rev C 31 Parker Hannifin 2015 www parker com hfde ...
Page 32: ...B 89 809 Rev C 32 Parker Hannifin 2015 www parker com hfde ...