Bulletin TI-N2-1100K 

Installation, Operation and Maintenance Manual

Quick Start Summary 

Parker Balston Model HPN2-1100, UHPN2-1100

www.balstonlabgasgenerators.com

14

1-800-343-4048

IEC 1010 Installation 

Category II

IEC 1010 Pollution 

Degree 2

CSA Certification Standard 

CAN/CSA 22.2 No. 1010-92

IEC Safety Standard 

IEC 1010-1:1990+ A1 1992+ A2:1995/ENG1010-1:1993

UL Safety Standard 

UL3101-1, First Edition

UL Listed to U.S. and Canadian Safety Standards
Nitrogen Purity 

99.9999% 99.9999%

O

2

 , H

2

O, CO

2, 

CO Concentration

 

< 1.0 ppm 

< 1.0 ppm

Hydrocarbon Concentration

 

N/A 

< 0.1 ppm

Ar Concentration

 0.9% 

0.9%

Inlet Connection

 

1/4" NPT (female) 

1/4" NPT (female)

Outlet Connection

 

1/8" NPT (female) 

1/8" NPT (female)

Compressed Air Consumption  

1.5 SCFM (42 lpm) 

1.5 SCFM (42 lpm)

Min/Max Inlet Air Pressure 

(4) 

60 psig/125 psig (4.1 barg/8.6 barg) 

60 psig/125 psig (4.1 barg/8.6 barg)

Recommended Inlet Air Temperature

 

78°F (26°C) 

78°F (26°C)

Altitude

 

2000 m 

2000 m

Max Inlet Air Dewpoint

 

60°F (15°C) 

60°F (15°C)

Ambient Operating Temperature 

 

60°F to 100°F (16°C  to 38°C) 

60°F to 100°F (16°C  to 38°C)

Electrical Requirements

 

120VAC / 230VAC, 50/60 Hz, 25 watts 

120VAC / 230VAC, 50/60 Hz, 575 watts

Dimensions 

12"w X 16"d X 35"h  

12"w X 16"d X 35"h  

 

 

 

 

(31cm X 41cm X 89cm) 

(31cm X 41cm X 89cm) 

Shipping Weight 

110 lbs. (50 kg) 

110 lbs. (50 kg) 

Fuse Type   (120 VAC Model)

 

1 amp, Type T, 250 V 

5 amp, Type T, 250 V 

 

 

 

           

         (230 VAC Model)

 

1 amp, Type T, 250 V 

3.15 amp, Type T, 250 V 

 

 

 

               

     (100 VAC Model)

 

1 amp, Type T, 250 V 

4 amp, Type T, 250 V

 

 HPN2-1100 

UHPN2-1100

System Specifications

Table 1 

                             

Nitrogen Outlet Pressure and Flow Specifications

  

 

    

 

Inlet Air Pressure 

Maximum N

2

 Flow 

Maximum N

2

 Pressure

  125 psig  

8.6 barg 

1.1 lpm 

85 psig  

5.9 barg

  110 psig  

7.6 barg 

1.0 lpm 

75 psig  

5.2 barg

  100 psig  

6.9 barg 

0.9 lpm 

65 psig  

4.5 barg

  90 psig  

6.2 barg 

0.8 lpm 

60 psig  

4.1 barg

  80 psig  

5.5 barg 

0.7 lpm 

50 psig  

3.4 barg

  70 psig  

4.8 barg 

0.6 lpm 

45 psig  

3.1 barg

  60 psig  

4.1 barg 

0.5 lpm 

35 psig  

2.4 barg

Notes
1

  The Parker Balston Nitrogen Generator 

was designed to meet the specifications 

listed when operated in a manner consis-

tent with the recommendations presented 

in this manual.  If these operating specifi

-

cations are not met, then inferior product 

performance and shortened product life 

may result.

2

  The nitrogen generator should be run 

continuously to optimize the performance 

and reliability of the unit.

4

  The Parker Balston Nitrogen Generator 

installation should be thoroughly checked 

for leaks after any service activity has been 

performed on the system in particular or its 

installation on the whole.  Any leak could 

significantly reduce the purity of the outlet 

nitrogen stream.

5

  The Parker Balston Nitrogen Generator will 

not operate with an inlet air pressure less 

than 60 psig (4.1 barg). A pressure regulator 

must be installed upstream from the gen-

erator to maintain an air supply pressure 

of 60 psig to 125 psig (4.1 barg to 8.6 

barg).

Optional Accessories

 P/N
Indicating Oxygen Trap  

72092

Prefilter, if needed

 2002N-1B1-DX

Pressure Regulator 

(upstream) 72-130-V883

Pressure Regulator

 (downstream) 

W-425-4032-000

Surge Supressor

 

75043 (North America)  75044 (Europe)

Hydrocarbon Scrubber

 76080

Summary of Contents for Ultra Pure HPN2-1100

Page 1: ...ntenance Manual Quick Start Summary Parker Balston Model HPN2 1100 UHPN2 1100 www balstonlabgasgenerators com 1 1 800 343 4048 Parker Balston Ultra Pure Nitrogen Generators Installation Operation and...

Page 2: ...16 C 38 C Electrical page 4 User selectable 120 VAC or 230 VAC 50 60 Hz Compressed Air page 4 Inlet air line 1 4 or larger Minimum Compressor capacity 1 5 scfm 45 lpm Air quality instrument quality se...

Page 3: ...pply valve and set the inlet pressure regulator see illustration on page 2 Turn the power switch upward to the On position The System Check light should blink every 12 seconds 3 The nitrogen generator...

Page 4: ...i trogen Generators are designed to provide high purity nitrogen gas for analytical instrumenta tion and other laboratory applications The HP UHP Nitrogen Generators are certified to the electrical re...

Page 5: ...alyst module used for hydrocarbon removal is a stainless steel vessel filled with catalyst and assembled with a cartridge heater This catalyst module oxidizes hydrocarbons in the compressed air supply...

Page 6: ...catalyst module heater Model UHPN2 1100 only or the electrical connections to it This indicator light is inoperative in the Model HPN2 1100 generator Inlet Pressure Display The inlet pressure display...

Page 7: ...to the compressor providing air to the generator The hydrocarbon catalyst module can also be contaminated by high concentrations of lead sulfur or phosphorous compounds heavy metals and long chain pol...

Page 8: ...are incorporated into the downstream equipment being used there is no need to install them separately These accessories must be appropriate for high purity applications Pressure Regulator downstream...

Page 9: ...ween 60 psig and 125 psig 4 1 barg to 8 6 barg During the 5 hours of warm up there will be no flow of nitrogen from the generator Figure 5 Manual Shut Off Valve Indicating Oxygen Trap If visible verif...

Page 10: ...the power returns If the power supply is interrupted for more than 1 hour the initial 5 hour start up cycle will be repeated when the power returns The flow of nitrogen from the Parker Balston Nitrog...

Page 11: ...been assembled into a maintenance kit MK7692 Maintenance Kit for Model HPN2 1100 and MK7694 Maintenance Kit for Model UHPN2 1100 If necessary the Nitrogen Generator may be wiped clean with a dry cloth...

Page 12: ...ion Operation and Maintenance Manual Quick Start Summary Parker Balston Model HPN2 1100 UHPN2 1100 www balstonlabgasgenerators com 12 1 800 343 4048 Figure 6 Maintenance Items PARTICULATE FILTER P N 7...

Page 13: ...egrity of the product use only fuses of the size and type detailed in the specifications section of this bulletin This equipment has main fuses in both neutral and phase lines Use care when servicing...

Page 14: ...1100 UHPN2 1100 System Specifications Table 1 Nitrogen Outlet Pressure and Flow Specifications Inlet Air Pressure Maximum N2 Flow Maximum N2 Pressure 125 psig 8 6 barg 1 1 lpm 85 psig 5 9 barg 110 ps...

Page 15: ...ations can result in unconsciousness without any warning symptoms such as dizziness fatigue etc Cautions WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin g...

Page 16: ...Symbols 3 CELL UHP Printed in U S A Bulletin TI N2 1100K 03 2017 2003 2017 Parker Hannifin Corporation Parker Manufacturing UK Ltd Gas Separation Filtration Division EMEA Dukesway Team Valley Trading...

Reviews: