background image

Bulletin TI-N2-1100K 

Installation, Operation and Maintenance Manual

Quick Start Summary 

Parker Balston Model HPN2-1100, UHPN2-1100

www.balstonlabgasgenerators.com

3

1-800-343-4048

Outlet Flow Operating 

Parameters

 Minimum 

Maximum 

Maximum

 

Inlet Pressure 

Outlet Flow 

Outlet Pressure

  120 psig (8.3 barg) 

1.1 lpm 

85 psig (5.9 barg)

  110 psig (7.6 barg) 

1.0 lpm 

75 psig (5.2 barg)

  100 psig (6.9 barg) 

0.9 lpm 

65 psig (4.5 barg)

 

90 psig (6.2 barg) 

0.8 lpm 

60 psig (4.1 barg)

 

80 psig (5.5 barg) 

0.7 lpm 

50 psig (3.4 barg)

 

70 psig (4.8 barg) 

0.6 lpm 

45 psig (3.1 barg)

 

60 psig (4.1 barg) 

0.5 lpm 

35 psig (2.4 barg)

Start-up:

Remove the service panel and open the 2-way valve (see illustration below).

Open the inlet air supply valve and set the inlet pressure regulator (see illustration on page 

2).  Turn the power switch upward to the "On" position.  The "System Check" light should 

blink every 12 seconds.

The nitrogen generator requires 5 hours to reach specified purity.  During this time, there 

will be no flow from the unit.  When the green "System Ready" light comes on, open the 

external outlet valve (see illustration on page 2).

Adjust the outlet flow according to the operating parameters listed below.  If the outlet flow 

exceeds the rated flow for a given inlet pressure, the concentration of impurities in the gas 

stream will increase. 

  

If you have any questions or concerns, please call the Technical Services Department at 

800-

343-4048 or email at [email protected].  

For locations outside North America, 

please contact your local representative.

Operation

(See technical information 

bulletin for details))

Summary of Contents for Ultra Pure HPN2-1100

Page 1: ...ntenance Manual Quick Start Summary Parker Balston Model HPN2 1100 UHPN2 1100 www balstonlabgasgenerators com 1 1 800 343 4048 Parker Balston Ultra Pure Nitrogen Generators Installation Operation and...

Page 2: ...16 C 38 C Electrical page 4 User selectable 120 VAC or 230 VAC 50 60 Hz Compressed Air page 4 Inlet air line 1 4 or larger Minimum Compressor capacity 1 5 scfm 45 lpm Air quality instrument quality se...

Page 3: ...pply valve and set the inlet pressure regulator see illustration on page 2 Turn the power switch upward to the On position The System Check light should blink every 12 seconds 3 The nitrogen generator...

Page 4: ...i trogen Generators are designed to provide high purity nitrogen gas for analytical instrumenta tion and other laboratory applications The HP UHP Nitrogen Generators are certified to the electrical re...

Page 5: ...alyst module used for hydrocarbon removal is a stainless steel vessel filled with catalyst and assembled with a cartridge heater This catalyst module oxidizes hydrocarbons in the compressed air supply...

Page 6: ...catalyst module heater Model UHPN2 1100 only or the electrical connections to it This indicator light is inoperative in the Model HPN2 1100 generator Inlet Pressure Display The inlet pressure display...

Page 7: ...to the compressor providing air to the generator The hydrocarbon catalyst module can also be contaminated by high concentrations of lead sulfur or phosphorous compounds heavy metals and long chain pol...

Page 8: ...are incorporated into the downstream equipment being used there is no need to install them separately These accessories must be appropriate for high purity applications Pressure Regulator downstream...

Page 9: ...ween 60 psig and 125 psig 4 1 barg to 8 6 barg During the 5 hours of warm up there will be no flow of nitrogen from the generator Figure 5 Manual Shut Off Valve Indicating Oxygen Trap If visible verif...

Page 10: ...the power returns If the power supply is interrupted for more than 1 hour the initial 5 hour start up cycle will be repeated when the power returns The flow of nitrogen from the Parker Balston Nitrog...

Page 11: ...been assembled into a maintenance kit MK7692 Maintenance Kit for Model HPN2 1100 and MK7694 Maintenance Kit for Model UHPN2 1100 If necessary the Nitrogen Generator may be wiped clean with a dry cloth...

Page 12: ...ion Operation and Maintenance Manual Quick Start Summary Parker Balston Model HPN2 1100 UHPN2 1100 www balstonlabgasgenerators com 12 1 800 343 4048 Figure 6 Maintenance Items PARTICULATE FILTER P N 7...

Page 13: ...egrity of the product use only fuses of the size and type detailed in the specifications section of this bulletin This equipment has main fuses in both neutral and phase lines Use care when servicing...

Page 14: ...1100 UHPN2 1100 System Specifications Table 1 Nitrogen Outlet Pressure and Flow Specifications Inlet Air Pressure Maximum N2 Flow Maximum N2 Pressure 125 psig 8 6 barg 1 1 lpm 85 psig 5 9 barg 110 ps...

Page 15: ...ations can result in unconsciousness without any warning symptoms such as dizziness fatigue etc Cautions WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin g...

Page 16: ...Symbols 3 CELL UHP Printed in U S A Bulletin TI N2 1100K 03 2017 2003 2017 Parker Hannifin Corporation Parker Manufacturing UK Ltd Gas Separation Filtration Division EMEA Dukesway Team Valley Trading...

Reviews: