Bulletin TI-N2-1100K
Installation, Operation and Maintenance Manual
Quick Start Summary
Parker Balston Model HPN2-1100, UHPN2-1100
www.balstonlabgasgenerators.com
3
1-800-343-4048
Outlet Flow Operating
Parameters
Minimum
Maximum
Maximum
Inlet Pressure
Outlet Flow
Outlet Pressure
120 psig (8.3 barg)
1.1 lpm
85 psig (5.9 barg)
110 psig (7.6 barg)
1.0 lpm
75 psig (5.2 barg)
100 psig (6.9 barg)
0.9 lpm
65 psig (4.5 barg)
90 psig (6.2 barg)
0.8 lpm
60 psig (4.1 barg)
80 psig (5.5 barg)
0.7 lpm
50 psig (3.4 barg)
70 psig (4.8 barg)
0.6 lpm
45 psig (3.1 barg)
60 psig (4.1 barg)
0.5 lpm
35 psig (2.4 barg)
Start-up:
1
Remove the service panel and open the 2-way valve (see illustration below).
2
Open the inlet air supply valve and set the inlet pressure regulator (see illustration on page
2). Turn the power switch upward to the "On" position. The "System Check" light should
blink every 12 seconds.
3
The nitrogen generator requires 5 hours to reach specified purity. During this time, there
will be no flow from the unit. When the green "System Ready" light comes on, open the
external outlet valve (see illustration on page 2).
4
Adjust the outlet flow according to the operating parameters listed below. If the outlet flow
exceeds the rated flow for a given inlet pressure, the concentration of impurities in the gas
stream will increase.
If you have any questions or concerns, please call the Technical Services Department at
800-
343-4048 or email at [email protected].
For locations outside North America,
please contact your local representative.
Operation
(See technical information
bulletin for details))