Bulletin TI-N2-1100K 

Installation, Operation and Maintenance Manual

Quick Start Summary 

Parker Balston Model HPN2-1100, UHPN2-1100

www.balstonlabgasgenerators.com

13

1-800-343-4048

 

      

Hydrocarbon

 

1st Stage 

2nd Stage 

3rd Stage 

Final Filter 

Catalyst

 

  

(76-94 only)

  

Module

Replacement P/N 

100-12-BX 

050-05-BX 

050-05-BX 

GS050-05-95 

76133 

Change Frequency 

Annual 

Annual 

Annual 

Annual 

3 years

Maintenance Kit - Model HPN2-1100 

MK7692

Maintenance Kit - Model UHPN2-1100 

MK7694

Ordering Information

Occasionally, one or both of the fuses in the UHP Nitrogen Generator may burn out.  The fuse 

located in the hydrocarbon removal module power harness is rated for 50 mA.  

This is not a field-

serviceable fuse and should not be removed under any circumstances. 

 The main fuses are located 

in the power receptacle on the side of the generator.  

Before servicing the main fuses, turn the 

generator off and disconnect the power cord from both the power supply and the generator power 

receptacle.

To access the main fuses, use a small screwdriver to remove the fuse holder located in the power 

receptacle of the generator.  In the UHP Nitrogen Generator, both the phase and neutral are fused 

separately.  As a result of these configuration, both fuses should be checked any time fuse replace-

ment is warranted

.  Replace either one or both fuses as necessary and re-assemble.  To maintain 

the safety and performance integrity of the product, use only fuses of the size and type detailed in 

the specifications section of this bulletin.

This equipment has main fuses in both neutral and phase lines.  Use care when servicing.

Fuse 

Replacement

1

2

3

The tools required to change the catalyst module are:  a Phillips head screwdriver, two 9/16" 

wrenches (or adjustable wrenches), and one 1/2" wrench (or adjustable).  The procedure for re-

placing the hydrocarbon catalyst module is as follows:

1

  Turn off the inlet air supply to the generator.  Wait approximately 10 minutes before turning 

the generator power off.  This will allow the generator to depressurize fully.

2

  Unplug the generator from the power receptacle.  

Allow the unit to cool for at least 2 hours 

before removing the cover.

3

  Remove the upper and lower service panels from the UHPN2-1100 (4 Phillips screws) and 

disconnect the electrical fittings to the hydrocarbon catalyst module at the connectors (see 

Figure 6).

4

  Disconnect the two tubing connections to the hydrocarbon catalyst module (see Figure 6) us-

ing the wrenches (one as an anchor).

5

  Remove the two screws which fasten the hydrocarbon catalyst module to the UHPN2-1100 

frame.

6

  Remove the spent module from the UHPN2-1100 and replace with the new module.

7

  Re-connect the inlet and outlet fittings to the hydrocarbon catalyst module, open the supply 

air valve, and check for leaks.  Re-connect the module's electrical connections.  Replace the 

service panels.  Plug the generator back into the wall receptacle.

8

  Restart the UHPN2-1100 by switching the power on.  The generator will begin its 5 hour 

warm-up.

Summary of Contents for Ultra Pure HPN2-1100

Page 1: ...ntenance Manual Quick Start Summary Parker Balston Model HPN2 1100 UHPN2 1100 www balstonlabgasgenerators com 1 1 800 343 4048 Parker Balston Ultra Pure Nitrogen Generators Installation Operation and...

Page 2: ...16 C 38 C Electrical page 4 User selectable 120 VAC or 230 VAC 50 60 Hz Compressed Air page 4 Inlet air line 1 4 or larger Minimum Compressor capacity 1 5 scfm 45 lpm Air quality instrument quality se...

Page 3: ...pply valve and set the inlet pressure regulator see illustration on page 2 Turn the power switch upward to the On position The System Check light should blink every 12 seconds 3 The nitrogen generator...

Page 4: ...i trogen Generators are designed to provide high purity nitrogen gas for analytical instrumenta tion and other laboratory applications The HP UHP Nitrogen Generators are certified to the electrical re...

Page 5: ...alyst module used for hydrocarbon removal is a stainless steel vessel filled with catalyst and assembled with a cartridge heater This catalyst module oxidizes hydrocarbons in the compressed air supply...

Page 6: ...catalyst module heater Model UHPN2 1100 only or the electrical connections to it This indicator light is inoperative in the Model HPN2 1100 generator Inlet Pressure Display The inlet pressure display...

Page 7: ...to the compressor providing air to the generator The hydrocarbon catalyst module can also be contaminated by high concentrations of lead sulfur or phosphorous compounds heavy metals and long chain pol...

Page 8: ...are incorporated into the downstream equipment being used there is no need to install them separately These accessories must be appropriate for high purity applications Pressure Regulator downstream...

Page 9: ...ween 60 psig and 125 psig 4 1 barg to 8 6 barg During the 5 hours of warm up there will be no flow of nitrogen from the generator Figure 5 Manual Shut Off Valve Indicating Oxygen Trap If visible verif...

Page 10: ...the power returns If the power supply is interrupted for more than 1 hour the initial 5 hour start up cycle will be repeated when the power returns The flow of nitrogen from the Parker Balston Nitrog...

Page 11: ...been assembled into a maintenance kit MK7692 Maintenance Kit for Model HPN2 1100 and MK7694 Maintenance Kit for Model UHPN2 1100 If necessary the Nitrogen Generator may be wiped clean with a dry cloth...

Page 12: ...ion Operation and Maintenance Manual Quick Start Summary Parker Balston Model HPN2 1100 UHPN2 1100 www balstonlabgasgenerators com 12 1 800 343 4048 Figure 6 Maintenance Items PARTICULATE FILTER P N 7...

Page 13: ...egrity of the product use only fuses of the size and type detailed in the specifications section of this bulletin This equipment has main fuses in both neutral and phase lines Use care when servicing...

Page 14: ...1100 UHPN2 1100 System Specifications Table 1 Nitrogen Outlet Pressure and Flow Specifications Inlet Air Pressure Maximum N2 Flow Maximum N2 Pressure 125 psig 8 6 barg 1 1 lpm 85 psig 5 9 barg 110 ps...

Page 15: ...ations can result in unconsciousness without any warning symptoms such as dizziness fatigue etc Cautions WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin g...

Page 16: ...Symbols 3 CELL UHP Printed in U S A Bulletin TI N2 1100K 03 2017 2003 2017 Parker Hannifin Corporation Parker Manufacturing UK Ltd Gas Separation Filtration Division EMEA Dukesway Team Valley Trading...

Reviews: