background image

Bulletin TI-N2-1100K 

Installation, Operation and Maintenance Manual

Quick Start Summary 

Parker Balston Model HPN2-1100, UHPN2-1100

www.balstonlabgasgenerators.com

11

1-800-343-4048

Allow at least 2 hours for the generator to cool prior to changing the catalyst module.

  The 

hydrocarbon catalyst module in the Parker Balston Nitrogen Generator is located behind the 

upper service panel and should be changed approximately every three years (or when the 

“Service System” LED starts blinking) to maintain the hydrocarbon removal specification for 

the unit.

Catalyst Module 

Maintenance

(Model UHPN2-1100 only)

Before servicing the Parker Balston Nitrogen Generator, isolate the unit from the com-

pressed air and electrical power supplies.  Allow the generator to cool for at least 2 hours 

before depressurizing (Model UHPN2-1100 only).  All maintenance tasks for the Parker 

Balston Nitrogen Generator should be performed by suitable personnel using reasonable 

care.

The primary maintenance tasks required by the Parker Balston Nitrogen Generator are  

changing the filter cartridges in the coalescing prefilter and particulate filters (annually), re-

placing the final membrane filter cartridge (annually), and replacing the hydrocarbon catalyst 

module (Model UHPN2-1100 only, 3 years).  A summary of the replacement part numbers and 

recommended service frequency is shown at the end of this Maintenance section.  See Figure 

6 for the locations of the various maintenance items.

Additional coalescing prefilter cartridges, particulate filter cartridges, and membrane filter 

cartridges for the Parker Balston HPN2-1100 and UHPN2-1100 Nitrogen Generators may 

be ordered through your local representative.  For convenience, a one-year supply of these 

replacement cartridges has been assembled into a maintenance kit (MK7692 Maintenance Kit 

for Model HPN2-1100 and MK7694 Maintenance Kit for Model UHPN2-1100).

If necessary, the Nitrogen Generator may be wiped clean with a dry cloth on an as needed 

basis.  

Do not use water, aerosols, or other cleaning agents to clean the unit.  The use of any 

liquid detergent to clean the generator could present an electrical hazard.

If the hydrocarbon catalyst module is not being replaced when the filters are replaced, there 

is no need to allow the generator to cool.  The only preparation needed to service the filters 

is turning off the air supply to the generator and depressurizing the affected filter hous-

ings.  To depressurize the filter housings, turn off the compressed air supply to the generator 

and allow the generator to cycle until the inlet pressure displays zero.  This should only take            

10 minutes (1-2 generator cycles).  Turn off power to generator.

The coalescing and particulate filters and the final membrane filter in the Nitrogen Generator 

are easily accessible when the lower service panel is removed (see Figure 6).  These filters 

should be changed on an annual basis to maintain efficient operation of the system.  The filter 

cartridge in the coalescing prefilter assembly is removed by loosening the bowl from the filter 

assembly, and unscrewing the element retainer from the base of the cartridge.  Insert the 

new filter cartridge (P/N 100-12-BX) and reassemble the housing in the reverse order.

The particulate filter(s) and final membrane filter cartridges are easily removed from the 

housing by unscrewing the filter bowl from the filter head (no tools required).  Insert the new 

filter cartridge (P/N 050-05-BX or GS050-05-95) and reassemble the filter housing.

Replace the lower service panel after replacing the filter cartridges.  The time required to 

service all stages of filtration integral to the generator is approximately 30 minutes.

Filter Cartridge 

Replacement

Maintenance

Summary of Contents for Ultra Pure HPN2-1100

Page 1: ...ntenance Manual Quick Start Summary Parker Balston Model HPN2 1100 UHPN2 1100 www balstonlabgasgenerators com 1 1 800 343 4048 Parker Balston Ultra Pure Nitrogen Generators Installation Operation and...

Page 2: ...16 C 38 C Electrical page 4 User selectable 120 VAC or 230 VAC 50 60 Hz Compressed Air page 4 Inlet air line 1 4 or larger Minimum Compressor capacity 1 5 scfm 45 lpm Air quality instrument quality se...

Page 3: ...pply valve and set the inlet pressure regulator see illustration on page 2 Turn the power switch upward to the On position The System Check light should blink every 12 seconds 3 The nitrogen generator...

Page 4: ...i trogen Generators are designed to provide high purity nitrogen gas for analytical instrumenta tion and other laboratory applications The HP UHP Nitrogen Generators are certified to the electrical re...

Page 5: ...alyst module used for hydrocarbon removal is a stainless steel vessel filled with catalyst and assembled with a cartridge heater This catalyst module oxidizes hydrocarbons in the compressed air supply...

Page 6: ...catalyst module heater Model UHPN2 1100 only or the electrical connections to it This indicator light is inoperative in the Model HPN2 1100 generator Inlet Pressure Display The inlet pressure display...

Page 7: ...to the compressor providing air to the generator The hydrocarbon catalyst module can also be contaminated by high concentrations of lead sulfur or phosphorous compounds heavy metals and long chain pol...

Page 8: ...are incorporated into the downstream equipment being used there is no need to install them separately These accessories must be appropriate for high purity applications Pressure Regulator downstream...

Page 9: ...ween 60 psig and 125 psig 4 1 barg to 8 6 barg During the 5 hours of warm up there will be no flow of nitrogen from the generator Figure 5 Manual Shut Off Valve Indicating Oxygen Trap If visible verif...

Page 10: ...the power returns If the power supply is interrupted for more than 1 hour the initial 5 hour start up cycle will be repeated when the power returns The flow of nitrogen from the Parker Balston Nitrog...

Page 11: ...been assembled into a maintenance kit MK7692 Maintenance Kit for Model HPN2 1100 and MK7694 Maintenance Kit for Model UHPN2 1100 If necessary the Nitrogen Generator may be wiped clean with a dry cloth...

Page 12: ...ion Operation and Maintenance Manual Quick Start Summary Parker Balston Model HPN2 1100 UHPN2 1100 www balstonlabgasgenerators com 12 1 800 343 4048 Figure 6 Maintenance Items PARTICULATE FILTER P N 7...

Page 13: ...egrity of the product use only fuses of the size and type detailed in the specifications section of this bulletin This equipment has main fuses in both neutral and phase lines Use care when servicing...

Page 14: ...1100 UHPN2 1100 System Specifications Table 1 Nitrogen Outlet Pressure and Flow Specifications Inlet Air Pressure Maximum N2 Flow Maximum N2 Pressure 125 psig 8 6 barg 1 1 lpm 85 psig 5 9 barg 110 ps...

Page 15: ...ations can result in unconsciousness without any warning symptoms such as dizziness fatigue etc Cautions WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin g...

Page 16: ...Symbols 3 CELL UHP Printed in U S A Bulletin TI N2 1100K 03 2017 2003 2017 Parker Hannifin Corporation Parker Manufacturing UK Ltd Gas Separation Filtration Division EMEA Dukesway Team Valley Trading...

Reviews: