background image

Parkell.com   |   

Questions?

 Contact Parkell Technical Support at (800) 243-7446 

Priming the Handpiece

1. 

Confirm the unit is ON and that the green LED on the front 
panel is illuminated.

2. 

Attach an autoclaved handpiece to the handpiece cable that has 
already been purged. Make sure there is no insert connected into 
the handpiece. (Purging the waterlines can only be accomplished 
when the handpiece is not connected to the cable.)

3. 

Hold the handpiece in an upright position.

4. 

Depress the foot pedal.

5. 

Water will begin to flow and fill the handpiece.  Priming of the 
handpiece should be done with a water setting between 
medium and maximum flow. (The lower the setting the 
longer it will take to prime and fill the handpiece with water.)

6. 

To stop water flow into the handpiece, remove foot from 
foot pedal.

7. 

An insert may now be placed into the handpiece for scaling.

Clinical Suggestions 

on Tooth Scaling

•  To prevent any potential injury to the patient, ultrasonic scalers 

should be used by trained, dental professionals only.

•  Protect patient’s eyes, lips, cheek, tongue or other vulnerable 

soft tissues when using this device. Inadvertent contact with 
Insert tip may cause a slight burn.

•  Inserts and detachable handpiece must be sterilized before 

each use.

•  Use of a face mask by the clinician, and high-volume intraoral 

suction, is strongly recommended when operating this device to 
avoid inhalation of contaminated aerosols.

•  To prevent a possible slight burn to the patient or clinician, 

NEVER OPERATE UNIT WITHOUT WATER FLOWING.

•  Use the lowest effective scaling power for the case at hand. This 

keeps heat generation to a minimum.

•  Before using the insert in the oral cavity, adjust the water spray 

following the graphics on the faceplate of the scaler over a sink 
or cuspidor until desired mist is obtained. Good water flow will 
act as a coolant and aid in flushing out debris to maximize 
patient comfort.  

•  If patient is new to ultrasonic scaling, explain to them what to 

expect. Usually, the patient should not experience discomfort. If 
patient experiences uncomfortable heat, adjust power and 
water controls accordingly.

•  Do not test a scaler tip for vibration on your fingers while the 

unit is operating. This is not a valid test of how scaling feels to 
the patient.

•  Hold the handpiece in a comfortable pen-grasp. To remove 

deposits from teeth, always use a light brushing stroke with the 
side of the insert tip, in contact with and parallel to the root surface. 
Excess pressure will not improve scaling action but may cause 
heating and pain to the patient. Keep the scaler insert in motion, 
using repeated gentle strokes to remove all tenacious deposits.

•  If the patient experiences discomfort due to the sound and 

vibration from the Scaling procedure, provide ear protection.

•  Do not use the point of the insert on the tooth surface. You may 

gouge the tooth.

•  Pause occasionally during scaling by removing foot from pedal 

to evaluate deposit removal.

•  Exercise caution near ceramic restorations, as they can discolor 

or fracture if stressed.

Common Operator Errors that May 

Result in Poor Clinical Performance

•  Failure to keep insert tip parallel to the long axis of the tooth.
•  Excessive use of hand pressure.
•  Failure to use the lowest effective power setting for the case 

at hand.

•  Insufficient water flow.
•  Scaling with the point of the insert instead of the side.
•  Inserts that are damaged, bent or worn out.

Cleaning and Infection Control

•  Consult www.CDC.gov for the “Guidelines for Infection Control in 

Dental Health - Care Settings” and “Statement on Reprocessing 
Dental Handpieces”.

•  DO NOT AUTOCLAVE the Turb control unit, as 

autoclave heat will destroy it. The control unit should be cleaned 
of debris with a water-damp paper towel, disinfected by wiping 
with a paper towel that has been saturated with an EPA-
registered low-level (HIV/HBV claim) to intermediate-level 
(tuberculocidal claim) hospital disinfectant, wiped of chemical 
residue with a water- damp paper towel, and dried. Utilize the 
surface disinfecting protocol of the disinfectant manufacturer.

•  Autoclaving does not remove debris that has accumulated on 

the handpiece or the insert. Failure to adequately remove debris 
will result in inadequate sterilization. Rinse the handpiece, and 
the insert under warm running water for 30 seconds to remove 
any external or internal soil or debris. Using a soft soapy cleaning 
brush to assist in the cleaning, if necessary. Use non-ammoniated 
detergent or dishwashing soap. Do not use ammoniated 
cleansers or disinfectants. Rinse the devices again under warm 
running water for 30 seconds to remove any residual soap and 
blot dry with a dry lint-free towel. Wipe Insert with a lint free 
towel that has been saturated in an EPA-approved, hospital-
grade intermediate or high-level disinfectant (minimum 2.5% 
Glutaraldehyde solution), following the instructions for use 
provided by the disinfectant manufacturer. Dry the devices with 
a dry lint-free towel.

•  The handpiece and inserts may be sterilized in any conventional 

steam autoclave following manufacturer’s instructions. A 
typical steam sterilization cycle is 132 +/- 2°C for 4 mins. 
(Vacuum) or 132 +/- 2°C for 15 mins. (Gravity), followed by a 
15-minute minimum cool-down period. Use a steam sterilization 
pouch that is compliant with ISO 1140-1 Type 4 and 11607 
standards. Once items are sterilized, adhere to the shelf-life 
specified by the pouch manufacturer. After sterilization, inspect 
the device in the autoclave bag for integrity. If suspicion about 
the item exists, discard it and order a replacement from Parkell 
or your dealer.

•  DO NOT USE DRY HEAT OR CHEMCLAVE on the handpiece or 

the insert.

•  The handpieces are designed to withstand a minimum of 1,000 

autoclaving cycles when reprocessed.

•  The ultrasonic inserts from Parkell are designed to withstand a 

minimum of 500 autoclaving cycles when reprocessed. 

Summary of Contents for TurboSensor +

Page 1: ...idered a substitute for formal training in ultrasonic scaling Appropriate dental professional education is strongly recommended prior to using this device clinically A01156revA0722 English Instructions for Use Parkell Technical Support Service 800 243 7446 M F from 8 30 AM to 5 00 PM EST ...

Page 2: ... of Power Cord 110V 12 6 380cm 220 230V 7 6 230cm Handpiece Operating Frequency 30 KHz hD722 Protection Against Electric Shock Class 1 Type B applied part Protection Against Ingress of Liquids Foot Switch Scaler IPX1 drip proof Power Supply IPX0 Ordinary Mode of Operation of Equipment Intermittent 10 minutes on 5 minutes off Operating Conditions 15 30 C 20 70 RH non condensing Transport and Storag...

Page 3: ... handpiece cable and the handpiece are clean and dry before assembling them for use As with all electrical devices the unit should not be immersed in fluids and fluids should not be allowed to enter the unit through front or rear panel slots Do not reach for the device if it has fallen into liquid until power is disconnected and do not use the device Return it to Parkell for servicing Modifying de...

Page 4: ...foot from pedal to evaluate deposit removal Exercise caution near ceramic or resin restorations as they can fracture if stressed Inserts Magnetostrictive Inserts for the TurboSensor scaler are available separately and are not included with the basic scaler unit The unit is designed to work with all Parkell and other market leading 30KHz inserts Brand new inserts may fit very tightly in the handpie...

Page 5: ...o enter Turbo mode A yellow indicator light will illuminate on the front panel to confirm the scaler is operating at Turbo power Scaling power to a point midway between the current setting and the scaler s maximum power Engaging the Turbo feature at a low power setting causes a significant boost in power Engaging the Turbo feature at high power will produce very little change Since water flow is n...

Page 6: ...ace You may gouge the tooth Pause occasionally during scaling by removing foot from pedal to evaluate deposit removal Exercise caution near ceramic restorations as they can discolor or fracture if stressed Common Operator Errors that May Result in Poor Clinical Performance Failure to keep insert tip parallel to the long axis of the tooth Excessive use of hand pressure Failure to use the lowest eff...

Page 7: ...logged Inspect and clear blockage or replace insert Water filter clogged Replace No water supply connected Correct Insert does not go into handpiece Insert stack is bent Straighten plates carefully by hand and reinsert O ring is too tight Lubricate O ring with water and use twisting motion to seat insert Detachable 30K Handpiece hD722 can only accept 30K insert Make sure that it is not a 25KHz ins...

Page 8: ...essories with the unit and include an explanation of the problem Pack in the original box add plenty of cushioning material and over box the unit during shipping Transit should maintain a dry temperature of 0 F 110 F You will be contacted for your approval of the repair along with any associated costs prior to any work The unit will be repaired and returned to you Outside the US repairs must be ma...

Reviews: