background image

Parkell.com   |   

Questions?

 Contact Parkell Technical Support at (800) 243-7446 

manufacturer. Otherwise, degradation of the performance of 
this equipment could result.

•  This equipment is not suitable for use in the presence of a 

flammable anesthetic gas mixture (when used along with air 
or oxygen). 

Clinical Precautions

•  Although the following precautions may be common sense to the 

experienced user of ultrasonic scalers, they bear repeating in the 
interest of safety, and the management of unnecessary risk.

•  Ultrasonic scalers should only be used by trained, dental 

professionals. 

•  Use of appropriate personal protection equipment and high-

volume intraoral suction is strongly recommended when operating 
this device, to avoid exposure to contaminated aerosols. 

• 

DO NOT OPERATE UNIT WITHOUT WATER FLOWING.

 Water 

flow through tip during use must always be enough to cool 
handpiece and insert. 

•  Do not allow prolonged contact of insert tip with lips, cheek, 

tongue, or other soft tissues. Insert tips are sharp and should be 
used carefully.

•  Handpiece and inserts supplied along with the unit are not 

sterile and needs to cleaned and sterilized in autoclave by the 
user prior to the first use and before each use, as specified on 
the insert package instructions and in Infection Control 
procedure specified in this instruction manual.

•  To keep heat generation to minimum, use the lowest scaling 

power that is effective. 

•  Before using the insert in the oral cavity, adjust water mist by 

varying the POWER & WATER settings. 

•  For patient comfort, use a generous flow of water as a coolant, 

lubricant, and debris flush. 

•  If patient is new to ultrasonic scaling, explain to them what to 

expect. When done properly, the patient should not experience 
any uncomfortable heat. If patient experiences excessive heat, 
adjust POWER and WATER controls accordingly, or examine 
inserts for wear or damage. 

•  Do not test a scaler tip on your fingers while operating, as this is 

not a valid test of how scaling feels to the patient. If you touch 
the tip, it will feel uncomfortably hot to your fingers. 

•  Hold the handpiece in a comfortable pen-grasp. Elevated heat 

may be felt if the lighted handpiece is gripped where the insert 
meets the handpiece.

•  Keep the long axis of the insert tip parallel to the long axis of the 

tooth to wipe deposits from the tooth with a light, brushing 
stroke with the side of the insert. Do not gouge the tooth with 
the point of the tip. 

•  Excess pressure will not improve scaling efficiency and may 

cause overheating of the tooth and pain to the patient. Keep the 
scaler insert in motion always, using repeated gentle strokes to 
remove all tenacious deposits. 

•  Occasionally pause during scaling by removing foot from pedal 

to evaluate deposit removal. Exercise caution near ceramic or 
resin restorations, as they can fracture if stressed.

Inserts

Magnetostrictive Inserts for the Turb scaler are available 
separately and are not included with the basic scaler unit. The unit 

is designed to work with all Parkell and other market-leading 30KHz 
inserts. Brand new inserts may fit very tightly in the handpiece 
initially and may be inserted and removed more easily by using a 
slight twisting motion, and by lubricating the insert “O” ring with 
water. Old, worn, or blunt inserts will perform poorly, generate 
excess heat, and should be replaced when necessary.

Installing Your Scaler

•  Parkell Turb System is designed to rest on a level 

surface. Be sure the unit is stable. This will assure that an 
accidental pull on the handpiece or corded foot pedal does not 
cause the unit to fall onto the floor.     

•  The scaler requires access to a source of drinking-quality water 

and a grounded electrical outlet.    

•  Locate the control unit such that it can be easily disconnected 

form the mains power line in case it needs to be disconnected 
for any reasons. Disconnection form the mains power line can 
be accomplished by removing the power cord from the wall 
plug, removing the power cord from the in-line “power supply 
adapter” or by removing the 2.5 mm metal plug connector from 
the back of the scaler.

•  Install the scaler where control panel and foot pedal will be easy 

to reach during scaling procedures.       

•  The rear of the unit requires access so the water filter may be 

changed periodically without difficulty.

•  The device and its separate power adapter generate a minimal 

amount of heat. Avoid covering them to allow normal cooling.

Water Connection

Before plugging in the device, connect the scaler’s water line to a 
drinking-quality water supply (15-35 psi optimal), free of sediment. 
An external, office-wide water filter is recommended to minimize 
frequent changes of the device’s in-line water filter. 

The male quick-connect that comes on the 
end of the water hose is the standard fitting 
presently used in the dental industry. If you 
are replacing a scaler that uses an Adec-
type or other type of connection, remove 
the old fitting from the unit hose and attach 
it to the new water line. Check for leaks at all 
hose and filter connections before use, and 
tighten if necessary.

Controlling the Water Flow

Observing the arrow printed on the 
control panel, turn the water control 
counter- clockwise to increase the flow, 
or clockwise to decrease water flow. If 
water does not flow through the insert 
when the foot pedal is depressed, the 
water passage in the scaling insert may 
be clogged.

Electrical Connection

The scaler must only be powered via the power supply assembly 
that is supplied with the device by Parkell. Power supply provided 
by Parkell is a Medical-Grade, IEC 60601 compliant, TUV approved 

Summary of Contents for TurboSensor +

Page 1: ...idered a substitute for formal training in ultrasonic scaling Appropriate dental professional education is strongly recommended prior to using this device clinically A01156revA0722 English Instructions for Use Parkell Technical Support Service 800 243 7446 M F from 8 30 AM to 5 00 PM EST ...

Page 2: ... of Power Cord 110V 12 6 380cm 220 230V 7 6 230cm Handpiece Operating Frequency 30 KHz hD722 Protection Against Electric Shock Class 1 Type B applied part Protection Against Ingress of Liquids Foot Switch Scaler IPX1 drip proof Power Supply IPX0 Ordinary Mode of Operation of Equipment Intermittent 10 minutes on 5 minutes off Operating Conditions 15 30 C 20 70 RH non condensing Transport and Storag...

Page 3: ... handpiece cable and the handpiece are clean and dry before assembling them for use As with all electrical devices the unit should not be immersed in fluids and fluids should not be allowed to enter the unit through front or rear panel slots Do not reach for the device if it has fallen into liquid until power is disconnected and do not use the device Return it to Parkell for servicing Modifying de...

Page 4: ...foot from pedal to evaluate deposit removal Exercise caution near ceramic or resin restorations as they can fracture if stressed Inserts Magnetostrictive Inserts for the TurboSensor scaler are available separately and are not included with the basic scaler unit The unit is designed to work with all Parkell and other market leading 30KHz inserts Brand new inserts may fit very tightly in the handpie...

Page 5: ...o enter Turbo mode A yellow indicator light will illuminate on the front panel to confirm the scaler is operating at Turbo power Scaling power to a point midway between the current setting and the scaler s maximum power Engaging the Turbo feature at a low power setting causes a significant boost in power Engaging the Turbo feature at high power will produce very little change Since water flow is n...

Page 6: ...ace You may gouge the tooth Pause occasionally during scaling by removing foot from pedal to evaluate deposit removal Exercise caution near ceramic restorations as they can discolor or fracture if stressed Common Operator Errors that May Result in Poor Clinical Performance Failure to keep insert tip parallel to the long axis of the tooth Excessive use of hand pressure Failure to use the lowest eff...

Page 7: ...logged Inspect and clear blockage or replace insert Water filter clogged Replace No water supply connected Correct Insert does not go into handpiece Insert stack is bent Straighten plates carefully by hand and reinsert O ring is too tight Lubricate O ring with water and use twisting motion to seat insert Detachable 30K Handpiece hD722 can only accept 30K insert Make sure that it is not a 25KHz ins...

Page 8: ...essories with the unit and include an explanation of the problem Pack in the original box add plenty of cushioning material and over box the unit during shipping Transit should maintain a dry temperature of 0 F 110 F You will be contacted for your approval of the repair along with any associated costs prior to any work The unit will be repaired and returned to you Outside the US repairs must be ma...

Reviews: