 Device  Description

The Turbo is a magnetostrictive ultrasonic scaler (D660) 
that operates at a frequency of 30KHz, which translates to the 
insert tip vibrating 30,000 times per second during active scaling. 
The Turb features a fully detachable and autoclavable 
handpiece for enhanced reprocessing and infection control 
between patients (D722). It also features an IntelliSense™ processor 
for ensuring top-level, consistent performance with all Parkell and 
market-leading 30k inserts. For more information about usage with 
non-Parkell inserts, please see Parkell’s warranty information.

The Turb also features: a tap-on “purge” mode for 
waterline cleansing; a “priming” mode for preparing the handpiece 
to begin scaling with the fi rst compression of the foot pedal; a 
“press and hold” turbo feature for assistance in removing 
challenging deposits if required; and a more precise water fl ow 
system for heightened aerosol control.

 Indications for Use

For removal of calculus, plaque, and oral debris from teeth during 
dental prophylaxis/scaling.

 Contraindications to Use

Because of the potential for electromagnetic interference, this 
device should not be used on patients or by clinicians with cardiac 
pacemakers, internal defi brillators, intracorporeal fl uid pumps or 
any other implantable electronic devices, or in close proximity to 
sensitive patient monitoring devices such as pulse oximeters.

If patient or operator is pregnant or has any medical condition 
which might be aff ected by this device during treatment, consult a 
physician prior to use.

In case of an adverse event, discontinue the scaling procedure.

Warnings

•  Close manual shut-off  valve on the dental offi  ce water supply 

every night before leaving the offi  ce.

•  The water / fl uid supply line to the scaler should always be 

turned off 

 whenever the device is being connected, 

disconnected, or when not in use.

•  Turn the scaler power off  every night before leaving the offi  ce 

with the rocker switch on the back. When power is off , the green 
LED on the front panel will turn off .

•  Locate the scaler and power adapter above and away from 

sources of fl uid that may enter them. Always ensure that the 
electrical connections on the handpiece cable and the handpiece 
are clean and dry before assembling them for use.

•  As with all electrical devices, the unit should not be immersed in 

fl uids, and fl uids should not be allowed to enter the unit through 
front or rear panel slots. Do not reach for the device if it has 
fallen into liquid until power is disconnected and do not use the 
device.  Return it to Parkell for servicing. Modifying device 
violates safety codes, endangers patients and operators, and 
voids the warranty.

•  Use of this equipment adjacent to or stacked with other 

equipment should be avoided.

•  Only use the provided power supply by the manufacturer along 

with the unit.

•  This equipment produces electromagnetic energy and may cause 

interference with other electronic devices. Should this occur, 
changing the position or location of the unit may be necessary. 

•  Use of accessories other than those specifi ed / provided by the 

manufacturer of this equipment could result in increased 
electromagnetic emissions, decreased electromagnetic 
immunity or improper operation.

•  The system is portable but must be handled with care when 

moving. Do not place the system on or next to a radiator or 
other heat source. Excessive heat may damage the system’s 
electronics. Place the system where air is free to circulate on all 
sides and beneath it.

•  Failure to follow properly validated sterilization processes and 

approved aseptic techniques for Parkell inserts or handpieces 
may result in cross contamination.

•  Insuffi  cient water fl ow could result in elevated water and the tip as 

well as insert stack temperature. When operated with suffi  cient 
water fl ow, the water and tip temperature should not exceed 50°C 
(122°F). Failure to follow recommendations could result in injury to 
patients or users. If temperature is elevated, increase water fl ow. If 
temperature remains elevated, discontinue use and contact Parkell.

•  Portable RF communications equipment (including peripherals 

such as antenna cables and external antennas) should be used 
no closer than 30 cm (12 inches) to any part of the [ME 
EQUIPMENT or ME SYSTEM], including cables specifi ed by the 

Detachable 

Autoclavable Handpiece.

Frequency: 30KHz

Water Control 

Knob

Purge Light—lights ON when 

PURGE mode is initiated.

Power Control 

Knob

ON Indicator Light—Lights 

ON when unit is turned on.

TURBO Light— Indicates the device is operating on 

high-power TURBO mode (footswitch fully depressed)

Power Adapter Connection

Main Power Control 

Rocker Switch

Water Connection

Summary of Contents for TurboSensor +

Page 1: ...idered a substitute for formal training in ultrasonic scaling Appropriate dental professional education is strongly recommended prior to using this device clinically A01156revA0722 English Instructions for Use Parkell Technical Support Service 800 243 7446 M F from 8 30 AM to 5 00 PM EST ...

Page 2: ... of Power Cord 110V 12 6 380cm 220 230V 7 6 230cm Handpiece Operating Frequency 30 KHz hD722 Protection Against Electric Shock Class 1 Type B applied part Protection Against Ingress of Liquids Foot Switch Scaler IPX1 drip proof Power Supply IPX0 Ordinary Mode of Operation of Equipment Intermittent 10 minutes on 5 minutes off Operating Conditions 15 30 C 20 70 RH non condensing Transport and Storag...

Page 3: ... handpiece cable and the handpiece are clean and dry before assembling them for use As with all electrical devices the unit should not be immersed in fluids and fluids should not be allowed to enter the unit through front or rear panel slots Do not reach for the device if it has fallen into liquid until power is disconnected and do not use the device Return it to Parkell for servicing Modifying de...

Page 4: ...foot from pedal to evaluate deposit removal Exercise caution near ceramic or resin restorations as they can fracture if stressed Inserts Magnetostrictive Inserts for the TurboSensor scaler are available separately and are not included with the basic scaler unit The unit is designed to work with all Parkell and other market leading 30KHz inserts Brand new inserts may fit very tightly in the handpie...

Page 5: ...o enter Turbo mode A yellow indicator light will illuminate on the front panel to confirm the scaler is operating at Turbo power Scaling power to a point midway between the current setting and the scaler s maximum power Engaging the Turbo feature at a low power setting causes a significant boost in power Engaging the Turbo feature at high power will produce very little change Since water flow is n...

Page 6: ...ace You may gouge the tooth Pause occasionally during scaling by removing foot from pedal to evaluate deposit removal Exercise caution near ceramic restorations as they can discolor or fracture if stressed Common Operator Errors that May Result in Poor Clinical Performance Failure to keep insert tip parallel to the long axis of the tooth Excessive use of hand pressure Failure to use the lowest eff...

Page 7: ...logged Inspect and clear blockage or replace insert Water filter clogged Replace No water supply connected Correct Insert does not go into handpiece Insert stack is bent Straighten plates carefully by hand and reinsert O ring is too tight Lubricate O ring with water and use twisting motion to seat insert Detachable 30K Handpiece hD722 can only accept 30K insert Make sure that it is not a 25KHz ins...

Page 8: ...essories with the unit and include an explanation of the problem Pack in the original box add plenty of cushioning material and over box the unit during shipping Transit should maintain a dry temperature of 0 F 110 F You will be contacted for your approval of the repair along with any associated costs prior to any work The unit will be repaired and returned to you Outside the US repairs must be ma...