background image

Parkell.com   |   

Questions?

 Contact Parkell Technical Support at (800) 243-7446 

User Maintenance and 

Authorized Service

The Parkell Turb requires very little maintenance for years 
of trouble-free operation. There are no user-serviceable parts 
inside the control unit or the foot pedal. Opening the cases will void 
all warranties. Authorized Parkell personnel must make all internal 
repairs. Parkell urges all scaler users to monitor the condition of the 
water filter periodically and change the filter whenever noticeable 
darkening of the disk occurs, or when water flow decreases. The 
most common reason for service for ultrasonic scalers is failure by 
the user to periodically change the water filter. This will result in 
contamination and clogging of the internal water components.

The ultrasonic inserts, autoclavable handpiece, power supply, water 
filter and water hose are all detachable and may be replaced by the 
user.  Contact Parkell for a list of authorized inserts and available 
replacement parts.

Within the US, all repairs must be made by Parkell. We maintain a 
complete service and parts facility in our factory at 300 Executive 
Drive, Edgewood, NY 11717. Equipment needing service in the US 
should be returned, freight pre-paid, via approved common carrier 
(e.g., USPS, UPS, FedEx), and adequately insured. Return all 
accessories with the unit and include an explanation of the problem. 
Pack in the original box, add plenty of cushioning material, and over 
box the unit during shipping. Transit should maintain a dry 
temperature of 0°F-110°F. You will be contacted for your approval 
of the repair, along with any associated costs, prior to any work. The 
unit will be repaired and returned to you.

Outside the US, repairs must be made by a Parkell-authorized facility.

If you have any questions or problems with the installation or use of 
your Turb, call Technical Support Service at 800-243-7446, 
M-F from 8:30 AM to 5:00 PM EST.

Warranty and Terms of Use

For full Warranty and Terms of Use information, please 
see www.parkell.com. Parkell’s Quality System is certified to 
ISO 13485.

Conformance to Standards

The Parkell Turb is TUV listed and conforms to IEC 60601-1, 
60601-1-2 (Electromagnetic Compatibility of Medical Devices) and 
CAN/ CSA C22.2 No. 601.1. Parkell’s quality system is certified to 
ISO13485.

Explanation of Symbols Used

Follow instructions for use

Package contents

Manufacturer

Qualified User Symbol

Medical Safety Classification Symbol

Temperature limitations

Do not use if package is damaged

Keep dry

Catalogue / stock number

Unique Device Identifier

Medical Device

Protective Earth Connection

Do not dispose this product into the ordinary 

municipal waste or garbage system

Single use only

Sterilizable in a steam sterilizer (autoclave) at the 

temperature specified

Footswitch Connection

AC Power Connection

Water Connection

Summary of Contents for TurboSensor +

Page 1: ...idered a substitute for formal training in ultrasonic scaling Appropriate dental professional education is strongly recommended prior to using this device clinically A01156revA0722 English Instructions for Use Parkell Technical Support Service 800 243 7446 M F from 8 30 AM to 5 00 PM EST ...

Page 2: ... of Power Cord 110V 12 6 380cm 220 230V 7 6 230cm Handpiece Operating Frequency 30 KHz hD722 Protection Against Electric Shock Class 1 Type B applied part Protection Against Ingress of Liquids Foot Switch Scaler IPX1 drip proof Power Supply IPX0 Ordinary Mode of Operation of Equipment Intermittent 10 minutes on 5 minutes off Operating Conditions 15 30 C 20 70 RH non condensing Transport and Storag...

Page 3: ... handpiece cable and the handpiece are clean and dry before assembling them for use As with all electrical devices the unit should not be immersed in fluids and fluids should not be allowed to enter the unit through front or rear panel slots Do not reach for the device if it has fallen into liquid until power is disconnected and do not use the device Return it to Parkell for servicing Modifying de...

Page 4: ...foot from pedal to evaluate deposit removal Exercise caution near ceramic or resin restorations as they can fracture if stressed Inserts Magnetostrictive Inserts for the TurboSensor scaler are available separately and are not included with the basic scaler unit The unit is designed to work with all Parkell and other market leading 30KHz inserts Brand new inserts may fit very tightly in the handpie...

Page 5: ...o enter Turbo mode A yellow indicator light will illuminate on the front panel to confirm the scaler is operating at Turbo power Scaling power to a point midway between the current setting and the scaler s maximum power Engaging the Turbo feature at a low power setting causes a significant boost in power Engaging the Turbo feature at high power will produce very little change Since water flow is n...

Page 6: ...ace You may gouge the tooth Pause occasionally during scaling by removing foot from pedal to evaluate deposit removal Exercise caution near ceramic restorations as they can discolor or fracture if stressed Common Operator Errors that May Result in Poor Clinical Performance Failure to keep insert tip parallel to the long axis of the tooth Excessive use of hand pressure Failure to use the lowest eff...

Page 7: ...logged Inspect and clear blockage or replace insert Water filter clogged Replace No water supply connected Correct Insert does not go into handpiece Insert stack is bent Straighten plates carefully by hand and reinsert O ring is too tight Lubricate O ring with water and use twisting motion to seat insert Detachable 30K Handpiece hD722 can only accept 30K insert Make sure that it is not a 25KHz ins...

Page 8: ...essories with the unit and include an explanation of the problem Pack in the original box add plenty of cushioning material and over box the unit during shipping Transit should maintain a dry temperature of 0 F 110 F You will be contacted for your approval of the repair along with any associated costs prior to any work The unit will be repaired and returned to you Outside the US repairs must be ma...

Reviews: