REPROCESSING
DANGER
Danger of electrocution
Liquids can conduct electricity, thereby posing a risk of elec-
tric shock.
• Always switch the compressor off and disconnect the power
plug from the mains socket before starting to clean it.
If the inhalation system is used
by only one patient
, follow the
instructions for cleaning and disinfecting in the section "Single
patient use" [see: Single patient use, page 26].
If the inhalation system is used
by several patients
, follow the
instructions for cleaning and disinfecting in the section "Multiple
patient use" [see: Multiple patient use, page 30].
If the inhalation system is used
in professional environ-
ments
, follow the information on reprocessing included in the
appendix at the end of these instructions for use.
4.1 Single patient use
Cleaning and disinfection cycles
Compressor housing
Clean when visibly dirty
Nebuliser (without con-
nection tubing) and mask
– Clean immediately after every use
– Disinfect at least once a day
Air filter
Replace after 200 operating hours
(approx. 1 year)
Nebuliser and mask
The mask can be cleaned, disinfected and dried together with
the nebuliser.
Preparation
MASK
• Detach the mask from the nebuliser.
• Detach the elastic band from the mask.
– 26 –
Summary of Contents for COMPACT2
Page 2: ......
Page 40: ...40...
Page 45: ...Limity o et ov n max 50 cykl o et ov n 45...
Page 46: ...46...
Page 50: ......
Page 98: ...50...
Page 102: ......
Page 145: ...Processing limits max 50 processing cycles 45...
Page 146: ...46...
Page 150: ......
Page 196: ...48...
Page 197: ...49...
Page 198: ...50...
Page 202: ......
Page 248: ...48...
Page 249: ...49...
Page 250: ...50...
Page 254: ......
Page 296: ...44...
Page 302: ...50...
Page 306: ......
Page 354: ...50...
Page 358: ......
Page 406: ...50...