Aerosol data according to ISO 27427
The aerosol characteristics presented in these instructions for
use were determined in accordance with ISO 27427 using Sal-
butamol. If other solutions or suspensions are used for
nebulisation, the aerosol characteristics may differ from the val-
ues shown (particularly if they have greater viscosity).
The following data is based on tests according to a standard
which takes adult breathing patterns as a basis. Therefore,
these figures will probably differ from corresponding figures
that were calculated for populations of children and infants.
Nozzle insert
(transparent)
Minimum com-
pressor flow
(3 l/min –
0.6 bar)
Nominal com-
pressor flow
(4 l/min –
1.2 bar)
7
Maximum com-
pressor flow
(6 l/min –
1.9 bar)
MMAD [µm]
8
5.4
4.5
3.4
GSD
9
2.13
2.23
2.31
Respirable fraction
[% < 5 µm]
44.0
53.2
66.2
Aerosol fraction
[% < 2 µm]
11.6
17.9
28.1
Aerosol fraction
[% > 2 µm < 5 µm]
32.4
35.3
38.1
Aerosol fraction
[% > 5 µm]
56.0
46.8
33.8
Aerosol output [ml] 0.30
0.40
0.38
Aerosol output rate
[ml/min]
0.09
0.12
0.14
Residual volume
[ml] (gravimetric)
1.28
1.15
1.14
7) Operation with PARI COMPACT2 compressor (Type 152).
8) MMAD = Mass Median Aerodynamic Diameter
9) GSD = Geometric Standard Deviation
– 37 –
Summary of Contents for COMPACT2
Page 2: ......
Page 40: ...40...
Page 45: ...Limity o et ov n max 50 cykl o et ov n 45...
Page 46: ...46...
Page 50: ......
Page 98: ...50...
Page 102: ......
Page 145: ...Processing limits max 50 processing cycles 45...
Page 146: ...46...
Page 150: ......
Page 196: ...48...
Page 197: ...49...
Page 198: ...50...
Page 202: ......
Page 248: ...48...
Page 249: ...49...
Page 250: ...50...
Page 254: ......
Page 296: ...44...
Page 302: ...50...
Page 306: ......
Page 354: ...50...
Page 358: ......
Page 406: ...50...