64
WARNING
n
Use the nurse call auxiliary button of this product as an auxiliary button,
and place the call button unit within the range where it can be operated
by the user. In the following cases, notification to the nurse call system
may not be possible even though the nurse call auxiliary button of this
product is pressed.
• When the power plug of the bed is not inserted into an outlet
• When power is not supplied to the bed due to a power failure or other
problem
• When the bed communication cable is not connected appropriately
• When the bed communication cable is damaged
Note
!
n
When the nurse call auxiliary button of this product will be used, be sure to check the op-
eration after connecting the bed communication cable.
• Check procedure
(1) Press the nurse call auxiliary button.
(2) Check that the LED lamp above the nurse call auxiliary button turns on and a nurse
call is made.
n
If the LED lamp above the nurse call auxiliary button does not turn on and the nurse call is
not made when the button is pressed, see “1. Troubleshooting” (p. 93 to 96) and then
check the bed and cable. If correct operation is not possible even after taking corrective
action, stop using the nurse call auxiliary button of this product and use the call button
unit.
n
For details on the suitable nurse call system and suitable bed communication cable, con-
tact Paramount Bed (see p. 116) or your distributor.