Foot end
Head end
57
X-ray cassette imaging (type with X-ray cassette tray only)
WARNING
n
Disconnect the power plug from the outlet and turn the battery off when
using X-ray equipment. The bed may malfunction resulting in injury or
damage.
n
Do not perform X-ray imaging using a cassette tray while an air mattress
or Nemuri SCAN is installed. The X-ray imaging results may be affected
by reflections of parts of an air mattress or Nemuri SCAN.
CAUTION
Use a cassette of the specified size. If other than the specified one is used,
the cassette may fall, resulting in injury or cassette damage or
deformation.
You can perform X-ray imaging using a cassette within the range of the diagonal lines in the figure below.
Supported cassette sizes
Size.................. Half: 38.8 x 46.4 cm or less Square: 38.8 x 38.8 cm or less
Thickness ........ 1.5 cm or less
Weight ............. 3.5 kg or less
(Unit: cm)
Note
!
n
When performing imaging by raising the head section, set the cassette all the way to the
back of cassette tray. In this case, the imaging range will be narrower than the range indi-
cated in the figure above.
n
X-ray imaging in an inclined state (Trendelenburg position) is not possible.
2
FUNCTIONS AND OPERATIONS
67
39
86
A seal providing an approximate indication of the
maximum imaging range is affixed to the head section
side.
When the cassette is set all the way to the back of the
cassette tray, the range up to the tip of the triangle on this
seal will be the imaging range (approximate indication).
Summary of Contents for A6 Series
Page 11: ...Foot end Foot end 10...
Page 118: ...117 Memo...
Page 119: ...118...
Page 120: ...Instruction manual edited by Torindo Co Ltd 2021 11...