104
Product name
Electrical Hospital Bed
Model No.*
7)
PA-H 6117
PA-H 6127
PA-H 6117 X
PA-H 6127 X
PA-H 6137
PA-H 6147
PA-H 6137 X
PA-H 6147 X
PA-H 6217
PA-H 6317
PA-H 6217 X
PA-H 6317 X
PA-H 6237
PA-H 6337
PA-H 6237 X
PA-H 6337 X
Electrical components
Type
Linear actuator (DC motor)
Power supply voltage/
frequency
100 to 240 VAC, 50/60 Hz
Power consumption
Head up: 120 W/240 VA or less
Knee up: 95 W/180 VA or less
Height adjustment: 270 W/495 VA or less
Maximum continuous
operating time
2 min. (downtime: 38 min.)*
8)
Rated USB output
5 V/1 A
Operating
45 dB or less*
9)
Fuse
Control box
250 VAC, 6.3 A, delay type, interrupting capacity: 100 A or more
Battery
specifications
Type/quantity
-
Small sealed lead acid battery
(12 VDC/1.2 Ah) x2
Rating
-
24 VDC/1.2 Ah (10 hour rate)
Materials
Handheld controller (accessory)
Front case: PPE/PS resin moldings, Rear case: ASA resin moldings,
Hook/Cable clip: POM resin molding, Cable: PVC resin molding
g
Frequently used functions
Head up, knee up, height adjustment, side-rail up/down operations
* IEC 60601-2-52: 2009+A1: 2015 application environments 1 and 2
*1) Beds with an “X” at the end of the part number come standard with an X-ray cassette tray.
*2) Bottom height indicates the height from the floor to the top of the bottom.
*3) Safe working load: Load that can safely use the bed.
(Weight of the user and the total weight of accessories such as mattresses and suitable products)
*4) Excluding model numbers PA-H 6127, PA-H 6127 X, PA-H 6147, and PA-H 6147 X.
Control panel is side-rail integrated panel and patient control panel.
*5) Model numbers PA-H 6127, PA-H 6127 X, PA-H 6147, and PA-H 6147 X only.
*6) When the shaking reaches a height of 43 cm from the floor to the top of the bottom, a warning sound is
heard and the shaking stops for safety reasons. Press the height button again to sweep to the lowest
height with a warning sound.
*7) The sixth number of the model number indicates the difference in the operation equipment.
• “1” and “3”: Side-rail integrated panel
• “2” and “4”: Handheld controller
*8) Time to protect electrical components from overheating when the bed is used continuously for maximum
continuous operating time.
*9) Based on the internal metrics.