PB-MF-MAN-046-REV1.0
Hot plate
14
The latency of the response is shown on the
screen to a precision of 0.01
seconds. The latency time remains on
screen until the pedal is pressed again
to run another test.
The procedure to be followed is described as follows:
8.2.1.
SETTING TO THE REQUIRED TEMPERATURE
1 Switch on the Hot Plate.
2 Press the SELECT button once. The present
required temperature will be displayed on
screen.
3 The decimal dot flashes, indicating that the value of the required temperature can
be modified.
4 Press the UP/DOWN ARROWS to modify the required
temperature.
5 Once the new required temperature has been reached, press the
SELECT button again.
6
This value will be stored permanently in the unit's memory and will prevail
throughout the work session.
7 There is no need to input the required temperature every day, as the value is
stored permanently in memory.
8 The led HEAT will remain on while the equipment is heating.
9 Following 5 minutes of pre-heating, the unit is ready for use.