PB-MF-MAN-046-REV1.0 

Hot plate 

 

13 

8.2.

 

START-UP 

When  the  instrument  is  switched  on  (switch  to  ON),  the 

  screen  displays  the 

present temperature of the hot plate (between 20ºC and 70ºC). If the temperature is 
above/below this, the screen will display: “––“. 
 

When the 

 button is pressed, the required temperature appears on 

 screen 

with a flashing decimal dot. 
 

Press  the 

 

  ARROWS  to  modify  (increase  or  reduce)  the  required 

temperature. Press the 

 button again to select the required temperature, which 

will be stored in the device's memory, eliminating the need to adjust it every day. 
 
While  the  required temperature  is  being  modified  the  plate  heating  system  remains 

active at the previously set temperature. The LED 

 remains on while the heating 

system is active. The amount of electrical power applied to the plate is proportional to 
the  difference  between  the  actual  and  the  target  plate  temperature.  Once  the 

required  temperature  is  reached,  the  led 

  will  go  off,  and  the  plate  will  stop 

heating. 
 
When  the  plate  temperature  falls  below  the  required  temperature  due  to  the  heat 
released into the environment, the heating system switches on automatically until the 
required temperature is reached again. 
 

WARNING

:  the  stability  of  the  plate's  temperature  does  not  only 

depend  on  the  heating  control  system,  but  also  on  environmental 
conditions. Stability is lower if the unit is in a place where ambient 
temperature  is  variable  (near  an  open  window,  air-conditioning 
unit, etc.) or if there is a substantial difference between the required 
temperature and the ambient temperature. 
 

Following  a  brief  pre-heating  period  (between  5  and  10  minutes,  depending  on  the 
difference  between  required  and  present  plate  temperature),  the  required 
temperature should be reached. The unit is now ready to begin testing with animals. 
 
The  test  consists  of  visually  detecting  a  characteristic  motor  response  in  mice  when 
they are placed on a hot plate, that is following a very brief period of exploration the 
mouse licks its leg. 
 
The user measures the time elapsed from when the animal is placed on the plate (start 
point: the timer is started by pressing the PEDAL once) until it licks its leg (end point: 
the timer is stopped by pressing the PEDAL again). 

Summary of Contents for LE7406

Page 1: ...PB MF MAN 046 REV1 0 Hardware User s Manual Hot plate Hot plate Panlab s l u C Energ a 112 08940 Cornell de Ll Barcelona Spain www panlab com International Calls 34 934 750 697 Domestic Call 934 190 7...

Page 2: ...caused directly or indirectly by the incorrect interpretation of what is expressed in the pages of this manual Some symbols may have more than one interpretation by professionals unaccustomed to thei...

Page 3: ...BORATORY PRACTICE Check all units periodically and after periods of storage to ensure they are still fit for purpose Investigate all failures which may indicate a need for service or repair Good labor...

Page 4: ...sport D Place the equipment on a flat surface and leave at least 10 cm of free space between the rear panel of the device and the wall Never place the equipment in zones with vibration or direct sunli...

Page 5: ...strain set up the PC display keyboard and mouse with correct ergonomic positioning according to your local safety guidelines WARNING Class A equipment is intended for use in an industrial environment...

Page 6: ...r or liquids on the equipment Once you have finished using the equipment turn it off with the main switch Clean and check the equipment so that it is in optimal condition for its next use The user is...

Page 7: ...ary Insert fuses in the fuse holder in the correct position CORRECT INCORRECT Figure 4 Fuses position 5 Insert the fuse holder again positioning it according to the voltage in the AC net 115V POSITON...

Page 8: ...SCRIPTION 10 7 1 FRONT PANEL 10 7 2 REAR PANEL 11 8 OPERATING THE LE 7406 12 8 1 EQUIPMENT CONNECTION 12 8 2 START UP 13 8 2 1 SETTING TO THE REQUIRED TEMPERATURE 14 8 2 2 TESTS WITH ANIMALS 15 8 3 EQ...

Page 9: ...by pressing a PEDAL when the animal is placed on the metal surface Following a period of latency the animal licks its leg This reaction is the usual evaluation criterion for the test The user presses...

Page 10: ...PB MF MAN 046 REV1 0 Hot plate 9 6 1 ACCESSORIES SUPPLIED LE 7406 HOT PLATE 1 PEDAL 1 Power supply cable 1 RS 232 connection cable 1...

Page 11: ...ROWS The Up arrow increases the setting temperature and the Down arrow lowers it The temperature can be set from 45 C to 62 C in increments of 0 1 C ARROWS These buttons are used to select the target...

Page 12: ...NEL Figure 8 Rear Panel PEDAL 6 35mm female jack The foot switch is connected to this connector RS 232 DB9 female connector to connect the control unit to the computer serial port POWER Power inlet ma...

Page 13: ...PMENT CONNECTION This illustration features a sample system connection Figure 9 Equipment connection The connections and necessary cables are listed in the following table FROM TO CABLE 1 LE7406 RS 23...

Page 14: ...falls below the required temperature due to the heat released into the environment the heating system switches on automatically until the required temperature is reached again WARNING the stability o...

Page 15: ...isplayed on screen 3 The decimal dot flashes indicating that the value of the required temperature can be modified 4 Press the UP DOWN ARROWS to modify the required temperature 5 Once the new required...

Page 16: ...EANING 8 3 1 CLEANING THE HOT PLATE To clean the hot plate you can use a cloth dampen with water and then dry it with a dry cloth In case that there would be traces of faeces and urine you can impregn...

Page 17: ...chapter 8 1 for instructions on how to connect this cable to the device If the computer does not have any serial port the RS232 USB adapter is needed ref CONRS232USB contact your local sales delegate...

Page 18: ...gram are correct The equipment will send data only when you press the pedal and the stopwatch is running then the time is stopped and is necessary to press the pedal again to start a new measurement T...

Page 19: ...PB MF MAN 046 REV1 0 Hot plate 18 11 PREVENTIVE MAINTENANCE EXPERIMENT HOT PLATE CLEANING TRANSPARENT CYLINDER CLEANING CHECK TRANSPARENT CYLINDER PLACING...

Page 20: ...switch 45 C 62 C 0 1 C 0 3 C 70 C without probe feedback 70 C TEMPERATURE PROBE Technology Measurement range showed Linearity Accuracy Active probe 20 C 70 C 0 1 C 0 1 C ENVIRONMENTAL CONDITIONS Opera...

Page 21: ...s WEEE Restricci n de ciertas Sustancias Peligrosas en aparatos el ctricos y electr nicos ROHS Restriction of the use of certain Hazardous Substances in electrical and electronic equipment ROHS Restri...

Page 22: ...PB MF MAN 046 REV1 0 Hot plate 21...

Reviews: