PAT1M / PAT1.5M OPERATION MANUAL
PAT1M/PAT1.5M: PA22222A01_03
Page: 19 of 36
Dec., 2007
ERROR 9 - Blocked or dirty exit sensor
Blocked or dirty exit sensor
Help
Air burst
Buzzer
Continue
POSSIBLE CAUSE:
This will occur if the exit (optical) sensor in the PDM Dispenser becomes obstructed due to a
cable tie or excess debris (dirt, grime, etc.) on the lens. The cable tie will remain in the dispenser or the end of the
PHM Feeder Hose that is attached to the dispenser. When this occurs, the tool cannot be actuated until the cable tie
is cleared from the system or hose.
CORRECTIVE ACTION:
Press the
"Buzzer"
button to turn the buzzer signal "off". Disconnect feeder hose from
dispenser. Look for cable tie in dispenser connector block and remove cable tie, if present. Reconnect feeder hose,
and resume operation. If the cable tie is not found, reconnect the feeder hose. Straighten the feeder hose to
eliminate any bends or twists and point the tool away from yourself and others. Press the
"Air burst"
button to
provide a burst of air to dislodge and advance the cable tie to the jaw area of the tool. The
"Air Burst"
button can be
pressed again if the cable tie does not advance to the jaw area. When the cable tie is advanced to the tool jaw area,
the "Tie in tool" message will appear and the buzzer will sound. Press the
"Buzzer"
button to turn the buzzer signal
"off". Remove the cable tie from the jaw area and then press the
"Continue"
button. Press the
"Load"
button to
load one cable tie into the dispenser and continue operation. (If
"Load"
does not appear, press the
"MENU"
button
to bring up the
"Reset"
display. Press the
"Reset load"
button to bring up the “Load, Unload, Menu” display.) If the
cable tie does not advance to the tool jaw area after depressing the
"Air Burst"
button several times, replace the
feeder hose with another and notify your maintenance department.
PREVENTION:
If this same error is repeated, the dispenser exit sensor lens may be cleaned with a cotton swab
dampened with isopropyl alcohol. Disconnect the feeder hose from the dispenser and insert the dampened end of
the cotton swab into the dispenser transfer tube about 1.5" (38 mm). Gently move the cotton swab in and out,
against the top and bottom walls of the transfer tube, to clean the sensor.
ERROR 10 - Blocked or dirty tool sensor
Blocked or dirty tool sensor
Help
Air burst
Buzzer
POSSIBLE CAUSE:
This will occur if the tool (optical) sensor in the tool becomes obstructed due to a cable tie or
excess debris (dirt, grime, etc.) on the lens. The cable tie will remain in the tool or the end of the PHM Feeder Hose
that is attached to the tool. When this occurs, the tool cannot be actuated until the cable tie is cleared from the tool or
feeder hose.
CORRECTIVE ACTION:
Press the
"Buzzer"
button to turn the buzzer signal "off". Disconnect feeder hose from
tool. Look for cable tie in tool or tool end of hose and remove cable tie, if present. Reconnect feeder hose, and
resume operation. If cable tie is not found, reconnect feeder hose. Straighten the feeder hose to eliminate any
bends or twists and point the tool away from yourself and others. Press the
"Air burst"
button to provide a burst of
air to dislodge and advance the cable tie to the jaw area of the tool. The
"Air Burst"
button can be pressed again if
the cable tie does not advance to the jaw area. When the cable tie is advanced to the tool jaw area, the "Tie in tool"
message will appear and the buzzer will sound. Press the
"Buzzer"
button to turn the buzzer signal "off". Remove
the cable tie from the jaw area and then press the
"Continue"
button. Press the
"Load"
button to load one cable tie
into the dispenser and continue operation. (If
"Load"
does not appear, press the
"Menu"
button to bring up the
"Reset"
display. Press the
"Reset load"
button to bring up the “Load, Unload, Menu” display.) If the cable tie does
not advance to the tool jaw area after depressing the
"Air Burst"
button several times, replace the feeder hose with
another and notify your maintenance department.
PREVENTION:
If this same error is repeated, the tool sensor lens may require cleaning. Contact your maintenance
department to have the tool sensor lens cleaned.
electronic components distributor