Product
Soak soft cloth with water or neutral detergent, squeeze the cloth
well and wipe off dirt.
(Standard frequency of cleaning: Once a month)
Do not use thinner, benzene, etc. that may give damage to the
product.
Exhaust tank/Exhaust tank at rear side
If the inside of the exhaust tank becomes dirty, pour hot water mixed
with neutral detergent into the tank and shake it well to remove dirt,
and rinse it thoroughly with water.
(Standard frequency of cleaning: Once a month)
Cover packings and chamber opening
Wipe them carefully with soft cloth impregnated water.
Contaminated cover packings and chamber opening may cause
steam leakage.
(Standard frequency of cleaning: Once a day)
(Standard frequency of cleaning: Once a day)
Chamber inside
・
・
Chamber inside
Wipe the chamber inside with soft cloth impregnated water or
neutral detergent. If it is difficult to get the dirt out of the chamber,
wipe it well with nylon scrubber impregnated with neutral detergent
and then rinse it with running water.
In the case of liquid sterilization, wash away the dirty liquid with
※
After sterilizing culture mediums that generated sulfide gas or
chlorine gas, clean the chamber on the inside without fail.
Corrosion of the chamber may cause premature malfunction
of the product.
water when it is boiling over.
(Standard frequency of cleaning: Once a month)
Bottom of chamber inside
Remove the heater cover and wash the bottom carefully by means
of a hand brush paying attention so as not to give damage to the
sensor.
(Standard frequency of cleaning: Once a month)
As for the cleaning of heavy soil, please consult your dealer.
Make sure
of power-OFF.
Temperature sensor
Drain port
Heater
(Plan of bottom of chamber inside)
Chamber opening
Cover packing
Maintenance
27
CAUTION
Set the power switch to
“
OFF
”
and sufficiently cool
the product before performing maintenance.
Failure to observe this warning may result in electric shock
or burns.
Before performing maintenance, set the power switch to the
OFF position.
・
・
・
・