4. Pour the heating water.
When using the attached stainless steel baskets in pile, be sure to
place the small basket on the top.
Set the baskets gently along the guides in the chamber.
Bring down the grip.
5. Place the sterilization objects inside.
1) Make sure on the digital display I that the temperature inside
the chamber is under 60
°
C
(
䇾㻌㻌㻌㻌㻌㻌䇿
appears when the temperature is under 25
°
C)
2) Check to see whether or not the cover packings and the
chamber opening are free from dirt or dust
If they are dirty or dusty, steam may leak to cause trouble. Keep
them always clean.
(Cover packings and chamber opening See page 27.)
3) Close the cover while bringing down the handle.
Then, a peep sound is heard, and the sterilization set temperature
appears on the digital display
㻌䊠㻌
and the sterilization time or the
sterilization time and cycle appear alternately on the digital display
䊡
.
6. Close the cover.
The amount of heating water is 3.5 liters.
Before, starting operation, check again how to place the sterilization
objects in reference to the
䇾
On Sterilization Objects
䇿
(page 13 ~ 14).
If the cover is closed when the temperature inside the chamber is over
60
°
C due to continuous operation, etc., the air inside the chamber will
expand suddenly to make it difficult to open and close the cover. In this
case, wait until the temperature will lower below
60 °C.
Purified water is best suited as heating water. If it is unavailable,
use city water instead.
MEMO
Advisory note
MEMO
MEMO
Water level
Tip (water level)
Heater cover
Heater cover
Heater
Heating water
Temperature
sensor
Hole
Handle
Peep
Make sure that the drain valve is closed.
Set the heater cover in the chamber.
Pour the water until the tip of the water level bracket of the heater
cover is immersed in the water. Since the amount of water decreases
with every sterilizing operation, check the water level ahead of
operation.
Small basket
Small basket
(Basket without bracket)
Bracket
Grip
Drain outlet
Small
Large
Small
Large
Large
Drain valve
How to Use
18
WARNING
ɿ
Do not block the holes or temperature sensor in the chamber
with objects.
Failure to observe this warning may destabilize control of
the product or result in incomplete sterilization.
ɿ
Do not place objects on the temperature sensor in the chamber.
(Continued)
ɿ
ɿ
ɿ
ɿ
ɿ
ɿ