
Be sure to wear thermally insulated leather gloves.
Note that it takes time until liquid is cooled down.
Handling a sterilized object with bare bands immediately
after the end of operation may result in burns.
Use care not to scald yourself when taking out the sterilized
objects at the end of operation.
Take out the sterilization objects when steam has been purged
completely from the chamber.
12.Take out the sterilization objects.
Set the power switch to OFF at the end of daily use.
13.Set the power switch to OFF.
To suspend the heat-retention, press the button. The heat-
retention is suspended. In the case of the melting/heat-retention
cycle, the cycle/program is set at this time automatically to the
1-1 program of the liquid sterilization cycle.
If the button is not pressed, the heat-retention continues for
72 hours.
For simple heat-retention, change the melting time to
䇾
0
䇿
in the
䇾
melting/heat-retention cycle
䇿㻌
and press the button.
When taking out the sterilization objects
(objects to be melted) during heat-retention.
For continuous use
When using the product consecutively, make sure on the digital display
䊠
th
at the temperature inside the chamber is below 60
°
C.
To check the current temperature inside the chamber in standby
status,press the and buttons simultaneously. (See page 15.)
If the cover is closed when the temperature inside the chamber is
raised due to continuous operation, etc., the air inside the chamber
will rapidly expand, making it difficult to open/close the cover.
Wait until the temperature lowers below 60
°
C.
Memo
14.Drain off the heating water.
1) Open the cover.
2) Open the drain valve.
3) After draining water, close the drain valve without fail.
Repeated use of the same heating water may corrode the chamber
and clog the piping.
Advisory note
2 hours later or longer after
the completion of operation
Drain hose
When the temperature inside
the chamber is 59
23
Drain off heating water everyday after use of the product.
Insert the attached drain hose in the drain water port
and drain water at least 2 hours after operation.
If water is drained out in haste, hot water may splatter, resulting
in burns or damage to the floor. Repeated use of the same
heating water may corrode the chamber or clog the piping.
CAUTION
CAUTION
ɿ
ɿ
ɿ
ɿ
ɿ
ɿ