2. Set the power switch to ON
The digital display of the control panel lights up. (The boiling point
setting temperature appears on the digital display
䊠
and
䇾
- - -
䇿
appears on the digital display
䊡
for the first 2 seconds.)
3. Open the cover.
(The cover can not be opened unless the power switch is turned ON)
Cover being CLOSE
Cover being OPEN
Settings
(
set temperature and
sterilization time/cycle) are indicated.
Digital display
䊠
: Sterilization temperature
Digital display
䊡
: Sterilization time and
cycle indicated alternately
This status is referred to as
䇾
Standby
Status
䇿
.
To check the temperature inside
the chamber,Press the
and buttons simultaneously.
The current temperature is indicated
while these buttons are kept pressed.
Digital display
䊠
: Temperature inside
the chamber
Digital display
䊡
: Temperature of
heating water
If the temperature is lower than 25
°
C,
䇾㻌㻌㻌㻌㻌㻌䇿㻌
appears.
The current temperature of the
heating water in the chamber is
indicated.
The figure shown above is for
the case when the temperature
of the heating water in the
chamber is 30
°
C.
1) Be sure to check the following points before opening
the cover.
If you attempt to open the cover by force, trouble may
be caused with the product.
Power switch is turned ON.
Pressure gauge indicates 0 MPa.
COVER LOCK lamp goes out.
4) Open the cover.
2) While pressing down the handle,
3) Pull the movable handle toward you.
The cover can not be unlocked unless the handle
is pressed down.
The figure shown above is for the
case when the boiling point set
temperature is 100
°
C.
Digital display
4)
Make sure that the power
switch is turned ON.
Make sure that the
COVER LOCK lamp goes out.
Movable
handle
Handle
2)
3)
Make sure that the pressure
gauge indicates
䇾
O
䇿
.
17