4
Display when stable sensing is possible.
Display when stable sensing is not possible.
1-point teaching in window comparator and hysteresis mode
With this method, you set the threshold range by setting the shift amount
to the desired value and measuring the incident light intensity once. The
shift amount will be added to the incident light intensity to provide the upper
threshold respectively subtracted from the incident light intensity to provide
the lower threshold.
*
For FX-502□, window comparator mode only works for sensing
output 1.
Window comparator mode
Hysteresis mode
1
1
1_SL
2_SL
P-1
1
1
1_SL
2_SL
P-1
1_SL
Lower threshold value
P‑1
Teaching point
2_SL
Upper threshold value
1
Shift amount. The shift amount is 10% by default and can be set in PRO
mode as a percentage or as the incident light intensity. If you select a
incident light intensity setting that exceeds the maximum or minimum, the
maximum or minimum sensitivity will be set automatically.
1. Press the SET key when no sensing object is
present.
2. Press the SET key when the sensing object is
present.
This sets the lower threshold 1_SL at 10%
(the set shift amount) lower than the incident
light intensity and the upper threshold 2_SL at
10% higher than the incident light intensity.
Display when stable sensing is possible.
Display when stable sensing is not possible.
2-point teaching in window comparator and hysteresis mode
With this method, you set the threshold range by performing teaching the
with two sensing objects with different incident light intensities (P‑1 and
P‑2).
*
For FX-502□, window comparator mode only works for sensing
output 1.
Window comparator mode
Hysteresis mode
1_SL (P-1)
2_SL (P-2)
1_SL (P-1)
2_SL (P-2)
1_SL (P‑1)
Teaching point 1 serving as the lower threshold value
2_SL (P‑2)
Teaching point 2 serving as the upper threshold value
1.
Press the SET key when the first sensing
object is present.
2. Press the SET key again when the second
sensing object is present.
Display when stable sensing is possible.
Display when stable sensing is not possible.
*
If you select a value exceeding the maximum/minimum, the
sensitivity will be set automatically to the maximum/minimum
value.
3-point teaching in window comparator and hysteresis mode
With this method, you set the threshold range by performing teaching with
three sensing objects with different incident light intensities (P‑1, P‑2, and
P‑3). After you have performed teaching, the three points P‑1, P‑2, and P‑3
will be sorted in ascending order into the values A, B, and C. The threshold
range will be calculated from A, B, and C as follows: The lower threshold
value 1_SL is the midpoint between A and B and the upper threshold value
2_SL is the midpoint between B and C.
*
For FX-502□, window comparator mode only works for sensing
output 1.
Window comparator mode
Hysteresis mode
1_SL
2_SL
A
B
C
1_SL
2_SL
A
B
C
A
Teaching point with the lowest incident light intensity
B
Teaching point with medium incident light intensity
C
Teaching point with the highest incident light intensity
1_SL
Lower threshold value (midpoint between A and B)
2_SL
Higher threshold value (midpoint between B and C)
1.
Press the SET key when the first sensing
object is present.
2. Press the SET key again when the second
sensing object is present.
3. Press the SET key again when the third sens‑
ing object is present.
Display when stable sensing is possible.
Display when stable sensing is not possible.
*
If you select a value exceeding the maximum/minimum, the
sensitivity will be set automatically to the maximum/minimum
value.