Model CTF878 Service Manual
1
Chapter 1.
1.1.4
Signal Setup
The Signal Setup option enables you to set parameters that affect the flowmeter’s signal processing and outputs:
• Zero Cutoff
• Velocity Averaging
• Errors Allowed
To enter the option, scroll to the Signal Setup entry on the Service Menu and press
[ENTER]
. The screen appears
similar to Figure 3 below.
Figure 3: Signal Parameter Window
1.1.4.1
Setting Up Signal Parameters
1.
Press the [
] arrow key to enter the window.
2.
The first prompt allows you to select or not select Reverse Flow.
a.
Press
[ENTER]
to select Reverse Flow, and press the [
] arrow key to enter the next prompt, or
b.
Press
[ENTER]
again to deselect Reverse Flow, and press the [
] arrow key to enter the next prompt.
AgcMgc CCorr Avg LP Filter
.5
Zero Cutoff
m/s
OK
Cancel
Velocity Averaging
Errors Allowed
2
None
Reverse Flow
Signal Setup
Signal
Summary of Contents for CTF878
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 6: ...Warranty iv Model CTF878 Service Manual no content intended for this page...
Page 8: ...Contents vi Model name Manual Type...
Page 18: ...Chapter 1 8 Model CTF878 Service Manual no content intended for this page...
Page 38: ...Chapter 3 28 Model CTF878 Service Manual no content intended for this page...
Page 40: ...Model CTF878 Service Manual 30...
Page 46: ...Chapter 4 36 Model CTF878 Service Manual no content intended for this page...
Page 54: ...Chapter 5 44 Model CTF878 Service Manual no content intended for this page...
Page 58: ...Appendix A 48 Doc Tittle Manual Type Date Description of Service Performed By...
Page 59: ...Doc Tittle Manual Type 49 Appendix A Date Description of Service Performed By...
Page 60: ...Appendix A 50 Doc Tittle Manual Type no content intended for this page...
Page 66: ......