5. Once installed, check that the appropriate LED on the UCI (labelled ‘SUPPORT’)
is illuminated, indicating that the powder port is correctly supported by the mixer
hardware. If the LED is not illuminated green at this stage, please refer to the
Troubleshooting table under section 7.
Figure 25
System inflated (regulated to 4 – 10 mbar) and support arm connected to the powder port,
energizing the ‘SUPPORT’ micro switch/interlock
Note: When the SHAFT, DOOR and SUPPORT interlocks are energized, the READY
light illuminates, indicating the system is ready to accept fluid and begin mixing.
Also the speed controller for the mixer will also be energized and will show the
word ‘STOP’ (See Figure 29).
16