4. Specifications
Table 1: Specifications
Footprint WxLxH:
40 x 114 x 89 cm (15.75 x 45 x 35 in.) - compact configuration
40 x 135 x 89 cm (15.75 x 53 x 35 in.) - extended configuration
Control box, cart, & mixer
enclosure material:
304 L stainless steel
Control box, Frame, Drive Box
surface finish
At least 1.2 μm Ra/ 47 μin Ra
Control box ingress rating:
IP 65
Drive Unit enclosure ingress
rating:
IP 23
Voltage:
100 – 230 VAC, 50/60 Hz,
Voltage fluctuation:
+/- 10%
Overvoltage Category:
II
Input Wattage:
Less than 250 W
Amperage:
100 V 2.2 A; 110 V 2.0 A; 230 V 1.0 A
Altitude rating:
1000 m
Operating environment:
Temperature 4-40°C, Max humidity 85%, avoid condensation
Pollution degree:
2 or better
Motor horsepower:
¼ hp
Power cord length:
600 cm (20 ft)
Power cord plug options:
US, Continental Europe, Swiss, Australia, Japan, UK
E-stop (present, yes/no, location): Yes, face of control box
Minimum & maximum speed:
20 to 300 RPM
Connectors for remote control:
TURCK RSFPV61; RSFPV579; HUMMEL M23 male thread,
12-pin.
Cord sets for remote control are not supplied
Functions available from remote
control panel:
Motor – start, stop
Speed – adjustment, indication
Alarm – indication
Mode of control (remote/local) – indication
Signal type(s) for remote
output/control:
Rotation speed out 4-20 mA, motor control in 0-10 V, discrete I/O
signals relay contact type
Alarms generated:
Motor failure, speed off-range failure, E-stop activation
Operator interface type:
Touchscreen PLC
Noise level at operator position:
63 dB
Casters:
2x swivel (front), 2X stationary (rear)
Wheel material:
Polyurethane
Recipe storage:
Yes. Up to ten can be stored. Up to ten instructions in each recipe
are executed sequentially. Each instruction contains individual
programmable parameters: mixing time, pause time, and speed.
Password protection:
Operator level: access to protected function - start recipe run.
Supervisor level: access to protected functions - Recipe Editor,
Pause or Abort Recipe run, Switch of Control between Local to
Remote, adjustment of set up for reduced set of parameters.
Maintenance level: access to protected functions – Program Setup
parameters, System parameters and includes Supervisor level of
access.