At the entry point dedicated screen is exposed while all controls (local and remote) as well as
motor rotation are blocked until local operator have ‘Reset to Remote’ button pressed (Figure 34).
That brings system to locked mode of operation functioning according to current settings on
remote panel.
Figure 33: Entry point reset
For details of remote control I/O signals, see section 15.
9.8. Alarms
Failure detected by the system will activate Failure mode while in any mode of operation. Failure
mode halts motor rotation, displays an alarm notification on screen (Figure 35 left) and generates
an alarm output signal for remote control. Alarm light changes the color from green to red and
audible signal comes to beeping.
The equipment in the drive unit will stay powered up but cannot be operated until failure reset is
complete. Failure resolution is available only locally from the magnetic mixer face panel. External
control equipment will only receive an alarm signal with no ability to feedback control.
Note: in locked mode remote panel will not receive alarm notification.
IMPORTANT:
If failure happens while any of editor screen is open the alarm is generated and
indicated via alarm light and sounder but notification will not appear on screen until escape from
editor to any of mode screens.
Figure 34: Alarm screen (left); Failure detection screen (right)
Reset of the failure mode is available from the failure detection screen (Figure 35 right) which is
opened after pressing the alarm button.
Upon alarm reset the system will return to one of the following:
a) Manual mode idle condition: if at the moment of alarm, the mixer is in remote mode or
idled in manual mode.