SECTION 4
OPERATION
53
4.5.2.1 TIME BASED TRANSIENT EXAMPLE
The following is an example of how to create a MIL-STD-704D time-based transient. (See section 4.4.2
regarding pre-stored transients)
Start with the previous program #4 example as shown in section 4.5.1. From any metering display press
PROGRAM, 4, ENTER, EDIT, EDIT,
and you should see the following display:
TRANSIENT: PRGM #4 SEGMENT #1 OF 1
T=0.0002 Va=108 Vb=108 Vc=108
F=400 WFa=1 WFb=1 WFc=1
Press
1, ENTER, ENTER,
to select segment 1 of 1.
TRANSIENT: PRGM #4 SEGMENT #1 OF 1
T=.0002 Va=108 Vb=108 Vc=108
F=400 WFa=1 WFb=1 WFc=1
Press
.OOO2, ENTER,
to select a time of .0002 seconds. This is the smallest time allowed for a time
based transient segment.
TRANSIENT: PRGM #4 SEGMENT #1 OF 1
T=0.0002 Va=108 Vb=108 Vc=108
F=400 WFa=1 WFb=1 WFc=1
Press
80, ENTER, 80, ENTER, 80, ENTER,
to select 80 volts on all three phases.
TRANSIENT: PRGM #4 SEGMENT #1 OF 1
T=0.0002 Va=80 Vb=80 Vc=80
F=400 WFa=1 WFb=1 WFc=1
Press
400, ENTER,
to select 400 Hz. output frequency.
TRANSIENT: PRGM #4 SEGMENT #1 OF 1
T=0.0002 Va=80 Vb=80 Vc=80
F=400 WFa=1 WFb=1 WFc=1