-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
▲
I
NSTRUCTION
M
ANUAL
TERAPRESS TP
-
7100S
Mounting Press
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1
-
520
-
882
-
6598 Fax +1
-
520
-
882
-
6599 email: [email protected] Web: https://www.metallographic.com
20
Table 4
-
1 Phenolic Mounting Powders
Table 4
-
2 Diallyl Phthalate Mounting Powders
* Grams of powder to produce a 5/8 inch mount without any sample
4.9 TP
-
7100S Mounting Guidelines
Table 4
-
3 Epoxy Mounting Powders
Phenolic Powders
1
-
inch
mounts
1.25
-
inch
mounts
1.5
-
inch
mounts
2
-
inch
mounts
Temperature (ºC)
160°C
(320°F)
160°C
(320°F)
160°C
(320°F)
160°C
(320°F)
Machine gage pressure (psi)
MPa
2150
14.8
2150
14.8
2150
14.8
2150
14.8
Time (minutes)
-
minimum
8
10
12
12
Grams of powder*
10.2
16.0
23
40.8
Diallyl Phthalate Powders
1
-
inch
mounts
1.25
-
inch
mounts
1.5
-
inch
mounts
2
-
inch
mounts
Temperature (ºC)
160°C
(320°F)
160°C
(320°F)
160°C
(320°F)
160°C
(320°F)
Machine gage pressure (psi)
MPa
2150
14.8
2150
14.8
2150
14.8
2150
14.8
Time (minutes)
-
minimum
8
10
12
12
Grams of powder*
10.2
16.0
23
40.8
Epoxy Powders
1
-
inch
mounts
1.25
-
inch
mounts
1.5
-
inch
mounts
2
-
inch
mounts
Temperature (ºC)
160°C
(320°F)
160°C
(320°F)
160°C
(320°F)
160°C
(320°F)
Machine gage pressure (psi)
MPa
2150
14.8
2150
14.8
2150
14.8
2150
14.8
Time (minutes)
-
minimum
8
10
12
12
Grams of powder*
13.7
21.4
30.8
55