I
NSTRUCTION
M
ANUAL
TERAPRESS TP
-
7100S
Mounting Press
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1
-
520
-
882
-
6598 Fax +1
-
520
-
882
-
6599 email: [email protected] Web: https://www.metallographic.com
2.4 Features
Mounting Force
The
TERAPRESS TP 7100S
applies force through a robust pneumatic system. Pressure is
precisely controlled via microprocessors.
Heating / cooling system efficiency
The heating elements and water
-
cooling jacket have been integrated into the upper part of the mold
cylinder for efficiency. Heating is controlled by a thermostat, with the heating times being variable
up to 30 minutes.
The cooling cycle begins automatically after the designated hold time at temperature has passed.
Cooling can be completed one of two ways:
1.
Cooling to a preset temperature
2.
Cooling for a specific amount of time
The
TERAPRESS TP 7100S
compression mounting press offers the user an abundance of control
for both the heating and cooling cycle allowing the
TP 7100S
the ability to mount any mounting
compound or sample composition.
6
Cooling Temperature (if
Mode is set to Temp)
Cooling Rate: Low or
High
Cooling Time (if Mode
is set to Time)
Cooling Mode: Temp
or Time
Current Temperature
Cooling Time Passed