Version 1.9 rev 21 Oct 2019
19
IKon-XL
Figure 5:
The CCU-500 Camera Control Unit (Rear View)
Power Input (& Power Switch)
There is a switched power input (I= on, O= Off) from the mains AC supply to the CCU-500. Note that there are fuses in
the line input (see Section 5.5 for details and replacement procedure).
Fans
3 fans provide cooling to the unit- ensure that there is sufficient space at the front and rear of the unit for unrestricted air
flow (minimum 100 mm).
2.4 p
oWeR
R
eqUiReMents
Ensure that the power connector for the iKon-XL camera is inserted securely. The connector is keyed to aid correct
orientation. Refer to technical specifications for power supply requirements.
NOTES:
1. The electrical mains lead should be certified for use in your country and in applicable countries the plug
must be fitted with a 240V 5A fuse.
2. The iKon-XL is designed to use the CCU-500 camera control unit for power and signal operations. If users
use any other power supply, they do so at their own risk.
Power Input
Fans (x3)