Version 1.9 rev 21 Oct 2019
18
IKon-XL
2.3 c
aMeRa
c
ontRol
U
nit
(ccU-500)
The Camera Control Unit provides control functions and a stable DC power supply for the iKon-XL.
CCU-500A
Supplied with the iKon-XL
Flexi-cooling Model
CCU-500B
Supplied with the iKon-XL
Deep Cooled Model
Figure 4:
The CCU-500 Camera Control Unit (front View)
Display Diagnostics
Switch display lights on or off. Switching the diagnostic lights off may be useful if background light levels must be
minimised.
Output Good
This LED (green) indicates that the power status of the CCU-500 is within normal parameters.
Thermal Alarm
This LED (red) illuminates to warn the user if the CCU-500 unit overheats e.g. due to obstruction to the cooling fans, or if
there is a fan failure.
Remote Shutdown
Enables remote power on/off using a cable with BNC connection. The LED (orange) illuminates if the unit has been
remotely powered down (Note there is no indication of power status in Solis).
•
CCU-500 On: Open-circuit (or logic 0).
•
CCU-500 Off: User must supply drive TTL logic high +2.4 V at 10 mA to powerdown the CCU-500.
Camera Cable Outlet
A camera control cable connects the CCU-500 to the iKon-XL camera. The standard supplied cable is 5 metres long- if
the CCU-500 is to be situated at a longer distance (15 m options available) this must be specified at time of ordering as
the CCU-500 must be configured for the cable used.
Warning
LEDs
Display
Diagnostics
Switch
Remote
Shut-down
Camera
Cable Outlet