4.Advanced User Guidebook
49
Mode
Holdoff
Auto
Normal
Single
Holdoff
Sensitivity
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then stop
100 ns - 10 s, adjust
M
knob or click
to set time
interval before another trigger occur, press
panel button or click
move cursor to choose
which digit to be set.
Set channel sensitivity(Range: 0.3div-10div).
Alternate Trigger (Trigger mode: Edge)
Trigger signal comes from two vertical channels when alternate trigger is on. This
mode is used to observe two unrelated signals. Trigger mode is edge trigger.
Alternate trigger (Trigger Type: Edge)
menu
list:
Menu
Settings
Instruction
Alternate
Mode
Edge
Set vertical channel trigger type as edge trigger.
Source
CH1
CH2
Channel 1 as trigger signal.
Channel 2 as trigger signal.
Coupling
AC
DC
Block the direct current component.
Allow all component pass.
Slope
Trigger on rising edge
Trigger on falling edge
Mode
Holdoff
Auto
Holdoff
Sensitivity
Acquire waveform even no trigger occurs
100 ns - 10 s, turn the
M
knob or click
to set
time interval before another trigger occur, press
panel button or click
to move
cursor to choose which digit to be set.
Set channel sensitivity(Range: 0.3div-10div).
Logic Trigger
Trigger according to logic relation.
In Logic Trigger mode, the trigger setting information is displayed on bottom right of
the screen, for example,
,
indicates that
trigger type is Logic, logic mode is AND, CH1 high level and trigger level is 0.00mV,
Summary of Contents for Dual-Channel Series
Page 1: ...XDS3000 Dual Channel Series Digital Storage Oscilloscopes User Manual www owon com ...
Page 29: ...3 Junior User Guidebook 22 Double Zoom Single Zoom ...
Page 42: ...4 Advanced User Guidebook 35 Figure 4 2 Rectangle window Figure 4 3 Blackman window ...
Page 43: ...4 Advanced User Guidebook 36 Figure 4 4 Hanning window Figure 4 5 Kaiser window ...