36
BIDI MODE
SETTING THE OPTICAL REFERENCE
Connect the MASTER and REMOTE together with the appropriate reference cables, as shown on the MASTER screen.
In this example, the input of the EF mode controller is connected to the MM port on each unit, and the output of the EF mode
controller is inserted into the OPM port.
Once the MASTER and REMOTE are connected as shown, press any key on the MASTER to continue.
The MASTER and REMOTE will begin communication, and the MASTER will show the progress of the communications.
REMOTE
TEST=Start testing fibers in link
DONE=Done testing fibers in link
SOURCE=Control the light source
035
MASTER
11
MM
OPM
MM
OPM
Encircled Flux
Mode Controller
Encircled Flux
Mode Controller
FIBER#1&2
Using your test reference
cables, please connect the master
and remote as shown and
press any key to proceed.
017
TEST
EXIT
MENU
SOURCE