Kind 2 Decimal point setup parameters (bP, SP, and CP)
In the parameter setup mode, a figure representing the decimal places of interest flickers.
MODE
・・・・・
Not used in the setup process.
RESET
・・・・
Each time the switch is turned ON, the decimal point moves one place to the left and the
figure increases by one.
→
When the decimal point appears at the desired location, hold MODE turned ON for 2 seconds. (This
establishes the setting and the window returns to review mode.
Example: Parameter "bP" (decimal place in instantaneous flowrate)
Relative to the figure at right of
decimal point, this point moves.
A figure representing the decimal places of interest (flickers)
Turning RESET button ON, the figure increases by one
If a setting above (bP = 2) is chosen, the instantaneous flowrate reads
[b1 □□□.□□]
.
(“0”
→
“1”
→
“2”
).
Kind 3 Analog output 4/20mA trim
◦
During the analog trim, keep on monitoring the analog output with a milliammeter or voltmeter.
Given below is the 4mA trim procedure (the same holds true with 20mA trim).
(1) At "A04 4.000," hold "MODE" turned ON for 2 seconds to go into the parameter setup mode; "0" blinks on
and off at the rightmost digit and a 4mA simulated output from the register appears.
(2) Set the milliammeter reading in the procedure below.
(If it reads 3.988mA, for example, set to "A04 3.988.")
Example: Parameter "A04" (4mA trim)
MODE
・・・
Each time the switch is turned ON, the digit of interest moves one place to the left.
RESET
・・・
Each time the switch is turned ON, the figure of interest increases by one.
Digit of interest (blinking)
• Turning MODE "ON," the blinking digit moves to the left.
• Turning RESET "ON," the figure increases by one ("0"
→
"1.")
·
Following entering the setting, hold MODE turned ON for 2 seconds to establish the setting.
(3) Analog output is trimmed: make sure of the milliammeter reading once again.
(The indicated reading returns to "A04 4.000" (the rightmost digit blinking) now.)
◦
If the reading is within the tolerance with respect to 4mA, by holding MODE turned ON for 2 seconds the
second time, you can exit the setup mode.
→
Adjustment is complete. Returns to the review mode.
◦
If the reading is still outside the tolerance, repeat step (2) above.