B-320-2-E
3
(7) Pull the internal assembly a little forward and
remove connectors CN1, CN2-1 (SV), and CN2-2
(MV) at the bottom of the battery pack.
CAUTION: With LW74E register,
CN2-2 (MV) is unused. At
reassembly, be careful
not to make a mistake in
connections.
CAUTION: Do not pull the sensor
bracket forcibly, or its
lead wires may break.
Now the internal assembly is ready for
removal.
INTERNAL ASSEMBLY
BATTERY PACK
(8) Remove the two posts securing the hold-down
strip with socket wrench.
Then remove three couplings at air connections
and silencer from their bushings.
CAUTION: Do not take a firm grip of
the silencer overmuch or
it will be crushed.
HEXAGONAL POSTS
COUPLING
BUSHING
SOLENOID
VALVE UNIT
HOLD DOWN STRIP
(9) Take off eight hex socket head bolts (M4) holding
the retainer, using hex wrench key.
CAUTION: At installation, tighten bolts
in an alternating order to
ensure even tightening.
RETAINER
(10) Draw the solenoid valve unit out of the register
body.
Now the solenoid valve is ready to be removed.
SOLENOID VALVE ASSEMBLIES
LW76E REGISTER SERVICE
(11) Loosen sensor retainer fitting screws (M4) with
screwdriver, remove the sensor retainer and
draw out the sensor unit.
The sensor unit is now ready to be removed.
SENSOR
RETAINER
SENSOR
CONNECTOR
SENSOR
RETAINER
FITTING
SCREW
(12) Insert a new sensor unit into the opening through
which the old sensor was removed. Assembly
is simply the reverse of the removal procedure.
Upon completion of assembly, perform an
operation check before service operation.
158 152
SENSOR ASSEMBLY
Location of Holes for Installing the Sensor
SENSOR FITTING
DISC
SENSOR OPENING
FOR SINGLE-CASE
METER
SENSOR OPENING
FOR DOUBLE-CASE
METER
(13) At register installation on the body, ensure that
the flow direction arrow on the body agrees with
the arrow mark on the sensor bracket.
FLOW DIRECION ARROW