6.23 Special Commands (SC) with EL4401
6.23.1 Function Codes 00 through 04
(
1
)
Description
This command allows you to batch read the settings in the parameter setup mode (data: liquid kind,
density, and density unit).
00: Batch read
01: Read liquid kind of liquid A 02: Read density and density unit of liquid A
03: Read liquid kind of liquid B 04: Read density and density unit of liquid B
(
2
)
Command/Response
Batch read
Example: Batch read
Code
Liquid kind
0
Crude
1
Fuel
2
Lube
* *
STX
STX
S C
F 0
F 0
ETX
CR LF
0 1
0 1
0 0
0 0 0 0
02 30 31 46 30 53 43 30 30 03
0D 0A
* *
ETX
CR LF
03
0D 0A
02 30 31 46 30 30 30
2B 31
30 30
30 30
30 30 35 43
30 30 30 2B
+ 1 0 0 0 0 0 + 0 0 5 C
0 0
30 30 2B 31 30 30
30 2B
30 30
30 30 30 30
+ 1 0 0 0 0 0 + 0 0 0 0
STX
STX
S C
ETX
BCC
CR LF
ETX
BCC
CR LF
Density
Density
Dens.
unit
HEX
HEX
Client
No.
Host
No.
Func.
code
Client
No.
Host
No.
Res-
ponse
Host system
PC, etc.
Client system
EL4401
Host system
PC, etc.
Client system
EL4401
Liquid
kind
Liquid
kind
Dens.
unit