L--740--28--E
7.10.2 Separately mounted sensor unit and transmitter wiring
Fig.7.9
Cut off shield wires here except for the
shield wire over BRN and RED lines.
Interconnect cable
Transmitter terminal box
Sensor terminal box
Shield
(Protected by black tube)
Shield
(Protected by black tube)
cut
cut
cut
Barrier cover to be applied
in a hazardous area
Brown
Red
Green
White
Blue
Grey
Purple
Yellow
Orange
Brown
Red
Green
White
Blue
Grey
Purple
Yellow
Orange
Interconnect cable (Max. 200m)
Sensor terminal box
シールド
(黒色チューブで保護)
カット
カット
カット
茶
赤
緑
白
青
灰
紫
黄
橙
茶
赤
緑
白
青
灰
紫
黄
橙
Shield wires:Black
Transmitter
TypeU
:
200m max.
TypeS Stainless steel tube type
:
50m max.
TypeS Titanium tube type
:
100m max.
TypeB
:
5m max.
CA00A, CA001, CA004
and CA006 to CA080
(T2
,
T3 and T4) are not
required.
NOTE 1. Do not fail to use dedicated interconnect cable.
2. Shield wire preparation
(1) Transmitter end:
As shown in the above figure, bundle shield wires colored in brown/red, green/white,
blue/grey, purple/yellow/orange, twist them, and cover the wires with a black tube. Then
connect only one wire to the terminal box (black) taking care to avoid potential contact
with the housing or conductive parts.
(2) Sensor end:
As shown in the figure, cover the brown/red shield wire with a black tube and connect it
to the terminal box taking care to avoid potential contact with the housing or conductive
parts. Clip all shield wires except brown/red as shown in the above figure.
3. Recommended cable end treatment:
Use of a stick type crimp terminal is not necessary.
9mm