Electrical fast transient/burst
IEC 61000-4-4
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines
Mains power quality should be that of a typical com-
mercial or hospital environment.
Surge IEC 61000-4-5
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
Mains power quality should be that of a typical com-
mercial or hospital environment.
Voltage dips, short inter-
ruptions and voltage vari-
ations on power supply input
lines IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
) for 0.5
cycle
40 % UT (60 % dip in U
T
) for 5
cycles
70 % U
T
(30 % dip in U
T
) for 25
cycles
<5 % U
T
(>95 % dip in U
T
) for 5 s
<5 % U
T
(>95 % dip in U
T
) for 0.5
cycle
40 % UT (60 % dip in U
T
) for 5
cycles
70 % U
T
(30 % dip in U
T
) for 25
cycles
<5 % U
T
(>95 % dip in U
T
) for 5 s
Mains power quality should be that of a typical com-
mercial or hospital environment. If the user of the
MADSEN Zodiac requires continued operation during
power mains interruptions, it is recommended that the
MADSEN Zodiac be powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels char-
acteristic of a typical location in a typical commercial or
hospital environment.
U
T
is the AC mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic immunity - for equipment and systems that are NOT life-supporting
MADSEN Zodiac is intended for use in the electromagnetic environment specified below. The user of MADSEN Zodiac should ensure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 V rms
150 kHz to 80 MHz
3 V rms
150 kHz to 80 MHz
Portable and mobile RF communications equipment
should be used no closer to any part of MADSEN Zodiac,
including cables, than the recommended separation dis-
tance calculated from the equation applicable to the fre-
quency of the transmitter.
Recommended separation distance:
d = 1.2
d = 1.2
for 80 MHz to 800 MHz
d = 2.3
for 80 MHz to 2.5 GHz,
where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation dis-
tance in metres (m).
Field strengths from fixed RF transmitters, as determ-
ined by an electromagnetic site survey,
a
should be less
than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment
marked with this symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
80 MHz to 2.5 GHz
35
User Guide
Otometrics - MADSEN Zodiac